Electrocardiographic recording direction impacts ventricular fibrillation waveform measurements: A potential pitfall for VF-waveform guided defibrillation protocols.
Thannhauser, Jos; Nas, Joris; Vart, Priya; Smeets, Joep L R M; de Boer, Menko-Jan; van Royen, Niels; Bonnes, Judith L; Brouwer, Marc A.
Afiliação
  • Thannhauser J; Department of Cardiology, Radboud University Medical Center, Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands.
  • Nas J; Department of Cardiology, Radboud University Medical Center, Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands.
  • Vart P; Department of Health Evidence, Radboud University Medical Center, Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands.
  • Smeets JLRM; Department of Cardiology, Radboud University Medical Center, Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands.
  • de Boer MJ; Department of Cardiology, Radboud University Medical Center, Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands.
  • van Royen N; Department of Cardiology, Radboud University Medical Center, Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands.
  • Bonnes JL; Department of Cardiology, Radboud University Medical Center, Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands.
  • Brouwer MA; Department of Cardiology, Radboud University Medical Center, Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands.
Resusc Plus ; 6: 100114, 2021 Jun.
Article em En | MEDLINE | ID: mdl-34223374
ABSTRACT

AIM:

In cardiac arrest, ventricular fibrillation (VF) waveform analysis has identified the amplitude spectrum area (AMSA) as a key predictor of defibrillation success and favorable neurologic survival. New resuscitation protocols are under investigation, where prompt defibrillation is restricted to cases with a high AMSA. Appreciating the variability of in-field pad placement, we aimed to assess the impact of recording direction on AMSA-values, and the inherent defibrillation advice.

METHODS:

Prospective VF-waveform study on 12-lead surface electrocardiograms (ECGs) obtained during defibrillation testing in ICD-recipients (2010-2017). AMSA-values (mVHz) of simultaneous VF-recordings were calculated and compared between all limb leads, with lead II as reference (proxy for in-field pad position). AMSA-differences between leads I and II were quantified using Bland-Altman analysis. Moreover, we investigated differences between these adjacent leads regarding classification into high (≥15.5), intermediate (6.5-15.5) or low (≤6.5) AMSA-values.

RESULTS:

In this cohort (n = 243), AMSA-values in lead II (10.2 ± 4.8) differed significantly from the other limb leads (I 8.0 ± 3.4; III 12.9 ± 5.6, both p < 0.001). The AMSA-value in lead I was, on average, 2.24 ± 4.3 lower than in lead II. Of the subjects with high AMSA-values in lead II, only 15% were classified as high if based on assessments of lead I. For intermediate and low AMSA-values, concordances were 66% and 72% respectively.

CONCLUSIONS:

ECG-recording direction markedly affects the result of VF-waveform analysis, with 20-30% lower AMSA-values in lead I than in lead II. Our data suggest that electrode positioning may significantly impact shock guidance by 'smart defibrillators', especially affecting the advice for prompt defibrillation.