Evaluation of a faecal calprotectin method using the OC-SENSOR PLEDIA.
O'Driscoll, Shane; Piggott, Carolyn; Benton, Sally C.
Afiliação
  • O'Driscoll S; NHS Bowel Cancer Screening Programme Southern Hub, Royal Surrey County Hospital, Guildford, UK.
  • Piggott C; NHS Bowel Cancer Screening Programme Southern Hub, Royal Surrey County Hospital, Guildford, UK.
  • Benton SC; NHS Bowel Cancer Screening Programme Southern Hub, Royal Surrey County Hospital, Guildford, UK.
Clin Chem Lab Med ; 60(6): 901-906, 2022 05 25.
Article em En | MEDLINE | ID: mdl-35285587
ABSTRACT

OBJECTIVES:

The National Institute for Health and Care Excellence recommends faecal calprotectin (f-cal) to help differentiate inflammatory bowel diseases from irritable bowel syndrome. Faecal samples for calprotectin have historically been collected at home by patients into screw-top pots and sent to laboratories where calprotectin is extracted and analysed. Faecal haemoglobin (f-Hb) samples are collected at home into specific collection devices containing stabilising buffer. We evaluated the OC-FCa method for f-cal, developed by Eiken Chemical Co., Ltd. (Japan) that uses the same collection device and analyser as f-Hb.

METHODS:

OC-FCa was assessed for limit of blank (LOB), limit of detection (LOD), limit of quantification (LOQ), within and between-run imprecision, linearity, prozone, recovery and carryover. A method comparison against the BÜHLMANN fCAL® turbo (BÜHLMANN Laboratories AG, Switzerland) was performed using patient samples and EQA.

RESULTS:

The LOB was 3 µg calprotectin/g faeces (µg/g), LOD 8 µg/g and LOQ 20 µg/g. Within and between-run imprecision was <5%; linearity was good (R2 > 0.99); prozone was appropriately detected; recovery was 99.6%; no observed carryover. OC-FCa showed a strong positive bias compared with BÜHLMANN fCAL® turbo (Z=-5.3587, p < 0.001). When categorised using our local pathway, which interprets calprotectin concentrations and need for further investigation, Cohen's Kappa demonstrates substantial agreement at <50 µg/g (κ=0.80) and >150 µg/g (κ=0.63) and fair agreement (κ=0.22) in the borderline category 50-150 µg/g.

CONCLUSIONS:

The OC-FCa method performed well in the evaluation. With the lack of standardisation for f-cal a clinical study is required to evaluate the positive bias and establish suitable cut-off levels.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doenças Inflamatórias Intestinais / Complexo Antígeno L1 Leucocitário Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Clin Chem Lab Med Assunto da revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Reino Unido

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Doenças Inflamatórias Intestinais / Complexo Antígeno L1 Leucocitário Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Clin Chem Lab Med Assunto da revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Reino Unido