Assessment and Comparison of Three Dimensional Exoscopes for Near-Infrared Fluorescence-Guided Surgery Using Second-Window Indocyanine-Green.
Cho, Steve S; Teng, Clare W; De Ravin, Emma; Singh, Yash B; Lee, John Y K.
Afiliação
  • Cho SS; Department of Neurosurgery, Hospital of the University of Pennsylvania, Philadelphia, PA, USA.
  • Teng CW; Perelman School of Medicine at the University of Pennsylvania, Philadelphia, PA, USA.
  • De Ravin E; Perelman School of Medicine at the University of Pennsylvania, Philadelphia, PA, USA.
  • Singh YB; Department of Neurosurgery, Hospital of the University of Pennsylvania, Philadelphia, PA, USA.
  • Lee JYK; Department of Neurosurgery, Hospital of the University of Pennsylvania, Philadelphia, PA, USA.
J Korean Neurosurg Soc ; 65(4): 572-581, 2022 Jul.
Article em En | MEDLINE | ID: mdl-35418003
ABSTRACT

OBJECTIVE:

Compared to microscopes, exoscopes have advantages in field-depth, ergonomics, and educational value. Exoscopes are especially well-poised for adaptation into fluorescence-guided surgery (FGS) due to their excitation source, light path, and image processing capabilities. We evaluated the feasibility of near-infrared FGS using a 3-dimensional (3D), 4 K exoscope with nearinfrared fluorescence imaging capability. We then compared it to the most sensitive, commercially-available near-infrared exoscope system (3D and 960 p). In-vitro and intraoperative comparisons were performed.

METHODS:

Serial dilutions of indocyanine-green (1-2000 µg/mL) were imaged with the 3D, 4 K Olympus Orbeye (system 1) and the 3D, 960 p VisionSense Iridium (system 2). Near-infrared sensitivity was calculated using signal-to-background ratios (SBRs). In addition, three patients with brain tumors were administered indocyanine-green and imaged with system 1, with two also imaged with system 2 for comparison.

RESULTS:

Systems 1 and 2 detected near-infrared fluorescence from indocyanine green concentrations of >250 µg/L and >31.3 µg/L, respectively. Intraoperatively, system 1 visualized strong near-infrared fluorescence from two, strongly gadoliniumenhancing meningiomas (SBR=2.4, 1.7). The high-resolution, bright images were sufficient for the surgeon to appreciate the underlying anatomy in the near-infrared mode. However, system 1 was not able to visualize fluorescence from a weakly-enhancing intraparenchymal metastasis. In contrast, system 2 successfully visualized both the meningioma and the metastasis but lacked high resolution stereopsis.

CONCLUSION:

Three-dimensional exoscope systems provide an alternative visualization platform for both standard microsurgery and near-infrared fluorescent guided surgery. However, when tumor fluorescence is weak (i.e., low fluorophore uptake, deep tumors), highly sensitive near-infrared visualization systems may be required.