Accelerated MRI at 9.4 T with electronically modulated time-varying receive sensitivities.
Glang, Felix; Nikulin, Anton V; Bause, Jonas; Heule, Rahel; Steffen, Theodor; Avdievich, Nikolai; Scheffler, Klaus.
Afiliação
  • Glang F; Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany.
  • Nikulin AV; Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany.
  • Bause J; Department of Biomedical Magnetic Resonance, Eberhard Karls University Tübingen, Tübingen, Germany.
  • Heule R; Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany.
  • Steffen T; Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany.
  • Avdievich N; Department of Biomedical Magnetic Resonance, Eberhard Karls University Tübingen, Tübingen, Germany.
  • Scheffler K; Magnetic Resonance Center, Max Planck Institute for Biological Cybernetics, Tübingen, Germany.
Magn Reson Med ; 88(2): 742-756, 2022 08.
Article em En | MEDLINE | ID: mdl-35452153
ABSTRACT

PURPOSE:

To investigate how electronically modulated time-varying receive sensitivities can improve parallel imaging reconstruction at ultra-high field.

METHODS:

Receive sensitivity modulation was achieved by introducing PIN diodes in the receive loops, which allow rapid switching of capacitances in both arms of each loop coil and by that alter B1- profiles, resulting in two distinct receive sensitivity configurations. A prototype 8-channel reconfigurable receive coil for human head imaging at 9.4T was built, and MR measurements were performed in both phantom and human subject. A modified SENSE reconstruction for time-varying sensitivities was formulated, and g-factor calculations were performed to investigate how modulation of receive sensitivity profiles during image encoding can improve parallel imaging reconstruction. The optimized modulation pattern was realized experimentally, and reconstructions with the time-varying sensitivities were compared with conventional static SENSE reconstructions.

RESULTS:

The g-factor calculations showed that fast modulation of receive sensitivities in the order of the ADC dwell time during k-space acquisition can improve parallel imaging performance, as this effectively makes spatial information of both configurations simultaneously available for image encoding. This was confirmed by in vivo measurements, for which lower reconstruction errors (SSIM = 0.81 for acceleration R = 4) and g-factors (max g = 2.4; R = 4) were observed for the case of rapidly switched sensitivities compared to conventional reconstruction with static sensitivities (SSIM = 0.74 and max g = 3.2; R = 4). As the method relies on the short RF wavelength at ultra-high field, it does not yield significant benefits at 3T and below.

CONCLUSIONS:

Time-varying receive sensitivities can be achieved by inserting PIN diodes in the receive loop coils, which allow modulation of B1- patterns. This offers an additional degree of freedom for image encoding, with the potential for improved parallel imaging performance at ultra-high field.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Imagem Assistida por Computador / Imageamento por Ressonância Magnética Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Alemanha