Your browser doesn't support javascript.
loading
Correlation of 3D Quantitative Coronary-Angiography Based Vessel FFR With Diastolic Pressure Ratio: A Single-Center Pooled Analysis of the FAST EXTEND and FAST II Studies.
Scoccia, Alessandra; Tomaniak, Mariusz; Neleman, Tara; Masdjedi, Kaneshka; Groenland, Frederik T W; Kardys, Isabella; Ligthart, Jurgen M R; Van Mieghem, Nicolas M; Spitzer, Ernest; Daemen, Joost.
Afiliação
  • Daemen J; Department of Cardiology, Room Rg- 628, Erasmus University Medical Center, P.O. Box 2040, 3000 CA Rotterdam, The Netherlands. j.daemen@erasmusmc.nl.
J Invasive Cardiol ; 34(9): E686-E688, 2022 09.
Article em En | MEDLINE | ID: mdl-35738562
ABSTRACT

BACKGROUND:

Vessel fractional flow reserve (vFFR) has a high diagnostic accuracy in assessing functional lesion significance compared with FFR. Nonhyperemic pressure ratios (NHPRs) were noninferior to FFR to guide revascularization of intermediate lesions. Therefore, the diagnostic performance of vFFR compared with NHPR warrants interest.

AIM:

To evaluate the diagnostic performance of vFFR with a generic diastolic pressure ratio (dPR) as a reference.

METHODS:

The study population was derived from the FAST EXTEND and FAST II studies. Between January 2016 and September 2020, a total of 475 patients were enrolled.

RESULTS:

Median dPR was 0.92 (interquartile range [IQR], 0.87-0.95), median vFFR was 0.86 (IQR, 0.80-0.90). The sensitivity, specificity, positive and negative predictive values, and diagnostic accuracy of vFFR ≤0.80 for dPR ≤0.89 were 66%, 92%, 79%, 85%, and 84%, respectively. Vessel FFR showed a good agreement with dPR (r=0.68), consistent among specific clinical lesion subsets and a high diagnostic accuracy for dPR ≤0.89 (area under the curve=0.89). Discordance between vFFR and dPR was observed in 78/492 cases (15.6%) and logistic regression analysis did not reveal any clinical, angiographic, or hemodynamic variables associated with vFFR and dPR discordance.

CONCLUSION:

Vessel FFR shows a good agreement with dPR and a high diagnostic accuracy for dPR ≤0.89.
Assuntos
Palavras-chave
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose Coronária / Reserva Fracionada de Fluxo Miocárdico Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: J Invasive Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2022 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Estenose Coronária / Reserva Fracionada de Fluxo Miocárdico Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: J Invasive Cardiol Assunto da revista: CARDIOLOGIA Ano de publicação: 2022 Tipo de documento: Article