Two Year Randomized Prospective Comparison of Ahmed Valve Versus Baerveldt Implant in Vitrectomized Eyes.
Kandarakis, Stylianos A; Petrou, Petros; Katsimpris, Andreas; Papakonstantinou, Evangelia; Timpilis, Marios; Chronopoulou, Konstantina; Lehman, Amanda; Ifantides, Cristos; Georgalas, Ilias.
Afiliação
  • Kandarakis SA; Department of Ophthalmology, National and Kapodistrian University of Athens, 1st University Eye Clinic, G. Gennimatas General Hospital, Athens, Greece.
  • Petrou P; Department of Ophthalmology, National and Kapodistrian University of Athens, 1st University Eye Clinic, G. Gennimatas General Hospital, Athens, Greece.
  • Katsimpris A; Department of Ophthalmology, National and Kapodistrian University of Athens, 1st University Eye Clinic, G. Gennimatas General Hospital, Athens, Greece.
  • Papakonstantinou E; Department of Ophthalmology, National and Kapodistrian University of Athens, 1st University Eye Clinic, G. Gennimatas General Hospital, Athens, Greece.
  • Timpilis M; Department of Ophthalmology, National and Kapodistrian University of Athens, 1st University Eye Clinic, G. Gennimatas General Hospital, Athens, Greece.
  • Chronopoulou K; Department of Ophthalmology, National and Kapodistrian University of Athens, 1st University Eye Clinic, G. Gennimatas General Hospital, Athens, Greece.
  • Lehman A; Scheie Eye Institute, University of Pennsylvania, Philadelphia, PA.
  • Ifantides C; Department of Ophthalmology, University of Colorado, Aurora, CO.
  • Georgalas I; Department of Surgery, Denver Health Medical Center, Denver, CO.
J Glaucoma ; 32(1): 27-33, 2023 01 01.
Article em En | MEDLINE | ID: mdl-36223288
ABSTRACT
PRCIS Ahmed Valve and Baerveldt shunt are efficacious options in vitrectomized eyes. Baerveldt implant achieves a lower mean intraocular pressure (IOP) at 2 years, with fewer medications and a higher percentage of medication-free patients.

PURPOSE:

To investigate and compare the efficacy and complications between Ahmed FP7 Glaucoma Valve (AGV) and Baerveldt 101-350 Glaucoma Implant (BGI) in vitrectomized eyes. MATERIALS AND

METHODS:

In this single-center randomized clinical trial, 43 vitrectomized eyes (39 patients) underwent glaucoma drainage device implantation. Eyes were randomized to receive either an AGV (FP7) or a BGI (101-350) and were followed for 2 years. Surgical success was defined as an IOP measurement≤18 mm Hg and≥5 mm Hg with or without glaucoma medication at 2 or more sequential visits after 3 months. The primary outcome was the comparison of the success rate at 2 years, while mean IOP, mean number of medications, and number of complications were considered secondary outcomes.

RESULTS:

Kaplan-Meier estimates of the 2-year success rates in IOP control after GDD implantation were similar between the 2 groups; AGV group 81.8% (95% CI 67.2%-99.6%) and BGI group 85.7% (95% CI 72.0%-100.0%), (log-rank test P value = 0.74). Patients in the BGI group had a statistically significant lower mean IOP compared with the AGV group in all follow-up visits at 2, 6, 12, and 24 months (11.62 vs. 17.45 mm Hg at the latter P value <0.001). The BGI group required a significantly lower number of medications for IOP control at the 2-year visit compared with the AGV group (0.76±0.99 vs. 1.5±1.06 P value = 0.02) but had a higher number of complications (62% vs. 41%, respectively).

CONCLUSIONS:

GDDs provide a viable solution for IOP control in vitrectomized eyes. Based on our prospective comparison, both Ahmed FP7 Glaucoma Valve and Baerveldt 101-350 Glaucoma Implant are efficacious options.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glaucoma / Implantes para Drenagem de Glaucoma Tipo de estudo: Clinical_trials Limite: Humans Idioma: En Revista: J Glaucoma Assunto da revista: OFTALMOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Grécia

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glaucoma / Implantes para Drenagem de Glaucoma Tipo de estudo: Clinical_trials Limite: Humans Idioma: En Revista: J Glaucoma Assunto da revista: OFTALMOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Grécia