Meta-Analysis of the Test-Retest Repeatability of [18F]-Fluorodeoxyglucose Standardized Uptake Values: Implications for Assessment of Tumor Response.
Shankar, Lalitha K; Huang, Erich; Litiere, Saskia; Hoekstra, Otto S; Schwartz, Larry; Collette, Sandra; Boellaard, Ronald; Bogaerts, Jan; Seymour, Lesley; deVries, Elisabeth G E.
Afiliação
  • Shankar LK; National Cancer Institute, Bethesda, Maryland.
  • Huang E; National Cancer Institute, Bethesda, Maryland.
  • Litiere S; European Organisation for Research and Treatment of Cancer Headquarters, Brussels, Belgium.
  • Hoekstra OS; Vrije Universiteit Medical Center, Amsterdam, the Netherlands.
  • Schwartz L; Columbia University Medical Center and New York Presbyterian Hospital, New York, New York.
  • Collette S; Bristol Myers Squibb, Princeton, New Jersey.
  • Boellaard R; Vrije Universiteit Medical Center, Amsterdam, the Netherlands.
  • Bogaerts J; European Organisation for Research and Treatment of Cancer Headquarters, Brussels, Belgium.
  • Seymour L; Queen's University, Kingston, Ontario, Canada.
  • deVries EGE; University of Groningen, Groningen, the Netherlands.
Clin Cancer Res ; 29(1): 143-153, 2023 01 04.
Article em En | MEDLINE | ID: mdl-36302172
ABSTRACT

PURPOSE:

Currently, guidelines for PET with 18F-fluorodeoxyglucose (FDG-PET) interpretation for assessment of therapy response in oncology primarily involve visual evaluation of FDG-PET/CT scans. However, quantitative measurements of the metabolic activity in tumors may be even more useful in evaluating response to treatment. Guidelines based on such measurements, including the European Organization for Research and Treatment of Cancer Criteria and PET Response Criteria in Solid Tumors, have been proposed. However, more rigorous analysis of response criteria based on FDG-PET measurements is needed to adopt regular use in practice. EXPERIMENTAL

DESIGN:

Well-defined boundaries of repeatability and reproducibility of quantitative measurements to discriminate noise from true signal changes are a needed initial step. An extension of the meta-analysis from de Langen and colleagues (2012) of the test-retest repeatability of quantitative FDG-PET measurements, including mean, maximum, and peak standardized uptake values (SUVmax, SUVmean, and SUVpeak, respectively), was performed. Data from 11 studies in the literature were used to estimate the relationship between the variance in test-retest measurements with uptake level and various study-level, patient-level, and lesion-level characteristics.

RESULTS:

Test-retest repeatability of percentage fluctuations for all three types of SUV measurement (max, mean, and peak) improved with higher FDG uptake levels. Repeatability in all three SUV measurements varied for different lesion locations. Worse repeatability in SUVmean was also associated with higher tumor volumes.

CONCLUSIONS:

On the basis of these results, recommendations regarding SUV measurements for assessing minimal detectable changes based on repeatability and reproducibility are proposed. These should be applied to differentiate between response categories for a future set of FDG-PET-based criteria that assess clinically significant changes in tumor response.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fluordesoxiglucose F18 / Neoplasias Tipo de estudo: Guideline / Systematic_reviews Limite: Humans Idioma: En Revista: Clin Cancer Res Assunto da revista: NEOPLASIAS Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fluordesoxiglucose F18 / Neoplasias Tipo de estudo: Guideline / Systematic_reviews Limite: Humans Idioma: En Revista: Clin Cancer Res Assunto da revista: NEOPLASIAS Ano de publicação: 2023 Tipo de documento: Article