Effects of different ablation settings on lesion dimensions in an ex vivo swine heart model: Baseline impedance, irrigant, and electrode configuration.
Jiang, Xi; Li, Sijie; Xiong, Qingsong; Zhang, Changzhi; Peng, Lijuan; Chen, Weijie; Cai, Yangwei; Yin, Yuehui; Chen, Shaojie; Ling, Zhiyu.
Afiliação
  • Jiang X; Department of Cardiology, the Second Affiliated Hospital of Chongqing Medical University, Chongqing, China.
  • Li S; Department of Cardiology, the Second Affiliated Hospital of Chongqing Medical University, Chongqing, China.
  • Xiong Q; Department of Cardiology, the Second Affiliated Hospital of Chongqing Medical University, Chongqing, China.
  • Zhang C; Department of Cardiology, the Second Affiliated Hospital of Chongqing Medical University, Chongqing, China.
  • Peng L; Department of Cardiology, the Second Affiliated Hospital of Chongqing Medical University, Chongqing, China.
  • Chen W; Department of Cardiology, the Second Affiliated Hospital of Chongqing Medical University, Chongqing, China.
  • Cai Y; Department of Cardiology, the Second Affiliated Hospital of Chongqing Medical University, Chongqing, China.
  • Yin Y; Department of Cardiology, the Second Affiliated Hospital of Chongqing Medical University, Chongqing, China.
  • Chen S; Department of Cardiology, the Second Affiliated Hospital of Chongqing Medical University, Chongqing, China.
  • Ling Z; Cardioangiologisches Centrum Bethanien (CCB), Frankfurt am Main, Germany.
J Cardiovasc Electrophysiol ; 34(1): 117-125, 2023 01.
Article em En | MEDLINE | ID: mdl-36403284
ABSTRACT

INTRODUCTION:

Intramural or epicardial locations of the arrhythmogenic substrate are regarded as one of the main reasons for radiofrequency (RF) catheter ablation failure. This study aims to conduct a comprehensive analysis of various factors including baseline impedance, irrigant and electrode configuration at similar ablation index (AI) value.

METHODS:

In 12 ex vivo swine hearts, RF ablation was performed at a target AI value of 500 and a multistep impedance load (100-180 Ω) in 4 settings (1) conventional unipolar configuration with an irrigant of normal saline (NS); (2) conventional unipolar configuration with an irrigant of half normal saline (HNS); (3) bipolar configuration with an irrigant of NS; (4) sequential unipolar configuration with an irrigant of NS. The relationships between lesion dimensions and above factors were examined.

RESULTS:

Baseline impedance had a strong negative linear correlation with lesion dimensions at a certain AI. The correlation coefficient between baseline impedance and depth, width, and volume were R = -0.890, R = -0.755 and R = -0.813, respectively (p < .01). There were 10 (total 10/100, 10%; bipolar 10/25, 40%) transmural lesions during the whole procedure. Bipolar ablation resulted in significantly deeper lesion than other electrode configurations. Other comparisons in our experiment did not achieve statistical significance.

CONCLUSION:

There is a strong negative linear correlation between baseline impedance and lesion dimensions at a certain AI value. Baseline impedance has an influence on the overall lesion dimensions among irrigated fluid and ablation configurations. Over a threshold impedance of 150 Ω, the predictive accuracy of AI can be compromised.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ablação por Cateter / Solução Salina Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Revista: J Cardiovasc Electrophysiol Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ablação por Cateter / Solução Salina Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Revista: J Cardiovasc Electrophysiol Assunto da revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: China