Bioinertization of NanoLC/MS/MS Systems by Depleting Metal Ions From the Mobile Phases for Phosphoproteomics.
Komori, Yumi; Niinae, Tomoya; Imami, Koshi; Yanagibayashi, Jun; Yasunaga, Kenichi; Imamura, Shinya; Tomita, Masami; Ishihama, Yasushi.
Afiliação
  • Komori Y; Graduate School of Pharmaceutical Sciences, Kyoto University, Kyoto, Japan.
  • Niinae T; Graduate School of Pharmaceutical Sciences, Kyoto University, Kyoto, Japan.
  • Imami K; Graduate School of Pharmaceutical Sciences, Kyoto University, Kyoto, Japan; Proteome Homeostasis Research Unit, RIKEN Center for Integrative Medical Sciences, Yokohama, Kanagawa, Japan.
  • Yanagibayashi J; Analytical & Measuring Instruments Division, Shimadzu Corporation, Kyoto, Japan.
  • Yasunaga K; Analytical & Measuring Instruments Division, Shimadzu Corporation, Kyoto, Japan.
  • Imamura S; Analytical & Measuring Instruments Division, Shimadzu Corporation, Kyoto, Japan.
  • Tomita M; Analytical & Measuring Instruments Division, Shimadzu Corporation, Kyoto, Japan.
  • Ishihama Y; Graduate School of Pharmaceutical Sciences, Kyoto University, Kyoto, Japan; Laboratory of Clinical and Analytical Chemistry, National Institute of Biomedical Innovation, Health and Nutrition, Ibaraki, Osaka, Japan. Electronic address: yishiham@pharm.kyoto-u.ac.jp.
Mol Cell Proteomics ; 22(5): 100535, 2023 05.
Article em En | MEDLINE | ID: mdl-36958626
ABSTRACT
We have successfully developed a bioinertized nanoflow LC/MS/MS (nanoLC/MS/MS) system for the highly sensitive analysis of phosphopeptides by depleting metal ions from the mobile phase. We found that not only direct contact of phosphopeptides with metal components, but also indirect contact with nanoLC pumps through the mobile phase causes significant losses during the recovery of phosphopeptides. Moreover, electrospray ionization was adversely affected by the mobile phase containing multiple metal ions as well as by the sample solvents contaminated with metal ions used in immobilized metal ion affinity chromatography for phosphopeptide enrichment. To solve these problems, metal ions were depleted by inserting an online metal ion removal device containing metal-chelating membranes between the gradient mixer and the autosampler. As a result, the peak areas of the identified phosphopeptides increased an average of 9.9-fold overall and 77-fold for multiply phosphorylated peptides with the insertion of the online metal ion removal system. This strategy would be applicable to the highly sensitive analysis of other phosphorylated biomolecules by microscale-LC/MS/MS.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fosfopeptídeos / Espectrometria de Massas em Tandem Idioma: En Revista: Mol Cell Proteomics Assunto da revista: BIOLOGIA MOLECULAR / BIOQUIMICA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fosfopeptídeos / Espectrometria de Massas em Tandem Idioma: En Revista: Mol Cell Proteomics Assunto da revista: BIOLOGIA MOLECULAR / BIOQUIMICA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Japão