Your browser doesn't support javascript.
loading
Comparison of 1.5 T and 3 T magnetic resonance angiography for detecting cerebral aneurysms using deep learning-based computer-assisted detection software.
Tajima, Taku; Akai, Hiroyuki; Yasaka, Koichiro; Kunimatsu, Akira; Yoshioka, Naoki; Akahane, Masaaki; Ohtomo, Kuni; Abe, Osamu; Kiryu, Shigeru.
Afiliação
  • Tajima T; Department of Radiology, International University of Health and Welfare Mita Hospital, 1-4-3 Mita, Minato-Ku, Tokyo, 108-8329, Japan.
  • Akai H; Department of Radiology, International University of Health and Welfare Narita Hospital, 852 Hatakeda, Narita, Chiba, 286-0124, Japan.
  • Yasaka K; Department of Radiology, International University of Health and Welfare Narita Hospital, 852 Hatakeda, Narita, Chiba, 286-0124, Japan.
  • Kunimatsu A; Department of Radiology, The Institute of Medical Science, The University of Tokyo, 4-6-1 Shirokanedai, Minato-Ku, Tokyo, 108-8639, Japan.
  • Yoshioka N; Department of Radiology, International University of Health and Welfare Narita Hospital, 852 Hatakeda, Narita, Chiba, 286-0124, Japan.
  • Akahane M; Department of Radiology, Graduate School of Medicine, The University of Tokyo, 7-3-1 Hongo, Bunkyo-Ku, Tokyo, 113-0033, Japan.
  • Ohtomo K; Department of Radiology, International University of Health and Welfare Mita Hospital, 1-4-3 Mita, Minato-Ku, Tokyo, 108-8329, Japan.
  • Abe O; Department of Radiology, International University of Health and Welfare Narita Hospital, 852 Hatakeda, Narita, Chiba, 286-0124, Japan.
  • Kiryu S; Department of Radiology, International University of Health and Welfare Narita Hospital, 852 Hatakeda, Narita, Chiba, 286-0124, Japan.
Neuroradiology ; 65(10): 1473-1482, 2023 Oct.
Article em En | MEDLINE | ID: mdl-37646791
ABSTRACT

PURPOSE:

To compare the diagnostic performance of 1.5 T versus 3 T magnetic resonance angiography (MRA) for detecting cerebral aneurysms with clinically available deep learning-based computer-assisted detection software (EIRL aneurysm® [EIRL_an]), which has been approved by the Japanese Pharmaceuticals and Medical Devices Agency. We also sought to analyze the causes of potential false positives.

METHODS:

In this single-center, retrospective study, we evaluated the MRA scans of 90 patients who underwent head MRA (1.5 T and 3 T in 45 patients each) in clinical practice. Overall, 51 patients had 70 aneurysms. We used MRI from a vendor not included in the dataset used to create the EIRL_an algorithm. Two radiologists determined the ground truth, the accuracy of the candidates noted by EIRL_an, and the causes of false positives. The sensitivity, number of false positives per case (FPs/case), and the causes of false positives were compared between 1.5 T and 3 T MRA. Pearson's χ2 test, Fisher's exact test, and the Mann‒Whitney U test were used for the statistical analyses as appropriate.

RESULTS:

The sensitivity was high for 1.5 T and 3 T MRA (0.875‒1), but the number of FPs/case was significantly higher with 3 T MRA (1.511 vs. 2.578, p < 0.001). The most common causes of false positives (descending order) were the origin/bifurcation of vessels/branches, flow-related artifacts, and atherosclerosis and were similar between 1.5 T and 3 T MRA.

CONCLUSION:

EIRL_an detected significantly more false-positive lesions with 3 T than with 1.5 T MRA in this external validation study. Our data may help physicians with limited experience with MRA to correctly diagnose aneurysms using EIRL_an.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aneurisma Intracraniano / Aprendizado Profundo Tipo de estudo: Diagnostic_studies / Observational_studies Limite: Humans Idioma: En Revista: Neuroradiology Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aneurisma Intracraniano / Aprendizado Profundo Tipo de estudo: Diagnostic_studies / Observational_studies Limite: Humans Idioma: En Revista: Neuroradiology Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Japão