Validation of a novel rebound tonometer (Tono-Vera® Vet) in normal ex vivo rabbit eyes.
Mills, Erinn P; Liu, Chin-Chi; Emelogu, Ugochi; Carter, Renee T; Camacho-Luna, Pilar; Lewin, Andrew C.
Afiliação
  • Mills EP; Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA.
  • Liu CC; Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA.
  • Emelogu U; Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA.
  • Carter RT; Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA.
  • Camacho-Luna P; Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA.
  • Lewin AC; Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA.
Vet Ophthalmol ; 26(6): 555-559, 2023 Nov.
Article em En | MEDLINE | ID: mdl-37658477
ABSTRACT

OBJECTIVE:

The purpose of this study was to validate the use of the Reichert Tono-Vera® Vet tonometer rabbit setting in normal ex vivo rabbit eyes and to compare the rabbit setting to the dog, cat, and horse settings of this tonometer. PROCEDURE Six freshly enucleated normal rabbit eyes were cannulated and connected to a fluid reservoir and physiologic monitor. Triplicate measurements were obtained with the four available settings dog, cat, horse, and rabbit at various intraocular pressures (IOP) ranging from 5 to 80 mmHg. Bland-Altman analysis was utilized to determine bias and 95% limits of agreement for each setting.

RESULTS:

Linear regression equations for the dog, horse, cat, and rabbit settings were y = 0.8101x + 2.5058, y = 0.7594x - 3.4673, y = 0.6635x + 0.3021, and y = 0.8935x + 1.3295, respectively. All settings demonstrated strong positive linear trends (dog r2 = 0.9644, horse r2 = 0.9456, cat r2 = 0.9309, and rabbit r2 = 0.9558). Bland-Altman plots revealed that the average bias and 95% limits of agreement (mmHg) were -4.73, -12.65, -12.86 and -2.73 and (-15.31, 5.86), (-29.03, 3.74), (-25.67, -0.05), and (-12.21, 6.76) for the dog, horse, cat, and rabbit settings, respectively.

CONCLUSION:

The Tono-Vera® Vet rabbit setting provided the most accurate and precise measurements compared with the other settings, but slightly underestimated actual IOP, especially as IOP was increased. This tonometer, using the rabbit setting, is likely to be appropriate for the estimation of IOP in rabbits with the appropriate correction formula applied.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tonometria Ocular / Pressão Intraocular Limite: Animals Idioma: En Revista: Vet Ophthalmol Assunto da revista: MEDICINA VETERINARIA / OFTALMOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tonometria Ocular / Pressão Intraocular Limite: Animals Idioma: En Revista: Vet Ophthalmol Assunto da revista: MEDICINA VETERINARIA / OFTALMOLOGIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Estados Unidos