Your browser doesn't support javascript.
loading
Safe and effective use of a passive safety needle by healthcare professionals in a simulated environment, including perceptions and preferences.
Serafin, Anna; Ryk, Aleksandra; Fendler, Wojciech.
Afiliação
  • Serafin A; Regulatory Affairs Department, HTL-STREFA S.A, Ozorkow, Poland.
  • Ryk A; Department of Biostatistics and Translational Medicine, Medical University of Lodz, Lodz, Poland.
  • Fendler W; Department of Biostatistics and Translational Medicine and Clinical Trials Unit, Medical University of Lodz, Lodz, Poland.
Expert Rev Med Devices ; 20(11): 963-971, 2023.
Article em En | MEDLINE | ID: mdl-37697473
ABSTRACT

BACKGROUND:

Needlestick injuries (NSIs) may potentially expose healthcare professionals (HCPs) to bloodborne pathogens. Safety needles are designed to protect against NSIs. We evaluated whether a new fully passive safety needle could be used safely by HCPs. RESEARCH DESIGN AND

METHODS:

The passive safety needle was tested by physicians, nurses, and pharmacists in subcutaneous or intramuscular injection scenarios in simulation studies (1-3). Data collected included successes, close calls, difficulties, use errors, and failures. In study 4, HCPs rated the device safety (21-item questionnaire).

RESULTS:

Overall, 104 participants completed 4772 simulated tasks, including 932 injections. 915 injections (98.18%) were performed successfully and no NSIs (0%) were observed in any of the studies. Studies 1 & 2 84.15% tasks and 96.06% injections were completed successfully, but use errors occurred, mostly arising from the participants' mental model. There were no failures in Study 3. In Study 4, >98% of participants responded positively to every question, while all felt that the passive safety feature could eliminate NSIs and would better protect against bloodborne pathogens than other existing devices with active or semi-passive safety mechanisms.

CONCLUSIONS:

The passive safety needle was used successfully by HCPs, did not lead to any NSIs, and was rated as the safest compared to similar devices.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Expert Rev Med Devices Assunto da revista: DIAGNOSTICO POR IMAGEM / TERAPEUTICA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Polônia

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Expert Rev Med Devices Assunto da revista: DIAGNOSTICO POR IMAGEM / TERAPEUTICA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Polônia