Your browser doesn't support javascript.
loading
Identifying Slim Modiolar Electrode Tip Fold-Over With Intracochlear Electrocochleography.
Varghese, Jordan J; Walia, Amit; Lefler, Shannon M; Ortmann, Amanda J; Shew, Matthew A; Durakovic, Nedim; Wick, Cameron C; Herzog, Jacques A; Buchman, Craig A.
Afiliação
  • Varghese JJ; Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA.
  • Walia A; Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA.
  • Lefler SM; Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA.
  • Ortmann AJ; Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA.
  • Shew MA; Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA.
  • Durakovic N; Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA.
  • Wick CC; Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA.
  • Herzog JA; Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA.
  • Buchman CA; Department of Otolaryngology-Head and Neck Surgery, Washington University School of Medicine, St. Louis, Missouri, USA.
Otolaryngol Head Neck Surg ; 170(4): 1124-1132, 2024 Apr.
Article em En | MEDLINE | ID: mdl-38018567
ABSTRACT

OBJECTIVE:

To evaluate the predictive value of intracochlear electrocochleography (ECochG) for identifying tip fold-over during cochlear implantation (CI) using the slim modiolar electrode (SME) array. STUDY

DESIGN:

Prospective cohort study.

SETTING:

Tertiary referral center.

METHODS:

From July 2022 to June 2023, 142 patients, including adults and children, underwent intracochlear ECochG monitoring during and after SME placement. Tone-bursts were presented from 250 Hz to 2 kHz at 108 to 114 dB HL. A fast Fourier transform (FFT) allowed for frequency-specific evaluation of ECochG response. ECochG patterns during insertion and postinsertion were evaluated using sensitivity and specificity analysis to predict tip fold-over. Intraoperative plain radiographs served as a reference standard.

RESULTS:

Fifteen tip fold-over cases occurred (10.6%) with significant ECochG response (>2 µV). Sixty-one cases without tip fold-over occurred (43.0%) with significant ECochG response. All tip fold-overs had both a nontonotopic postinsertion sweep and nonrobust active insertion pattern. No patients with robust insertion or tonotopic sweep patterns had tip fold-over. Sensitivity of detecting tip fold-over when having both nonrobust insertion and nontonotopic sweep patterns was 100% (95% confidence inteval [CI] 78.2%-100%), specificity was 68.9% (95% CI 55.7%-80.1%), and the overall accuracy was 72.0% (95% CI 60.5%-81.7%).

CONCLUSION:

Intracochlear ECochG monitoring during cochlear implantation with the SME can be a valuable tool for identifying properly positioned electrode arrays. In cases where ECochG patterns are nonrobust on insertion and nontonotopic for electrode sweeps, there may be a concern for tip fold-over, and intraoperative imaging is necessary to confirm proper insertion.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Implantes Cocleares / Implante Coclear Limite: Adult / Child / Humans Idioma: En Revista: Otolaryngol Head Neck Surg Assunto da revista: OTORRINOLARINGOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Implantes Cocleares / Implante Coclear Limite: Adult / Child / Humans Idioma: En Revista: Otolaryngol Head Neck Surg Assunto da revista: OTORRINOLARINGOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos