Implementation of the multielectrode radiofrequency-balloon in real-world clinical practice-operator learning curve and procedural outcome at a high-volume center.
Zylla, Maura M; Starrach, Lydia; Rahm, Ann-Kathrin; Thomas, Dierk; Frey, Norbert; Lugenbiel, Patrick.
Afiliação
  • Zylla MM; Department of Cardiology, Medical University Hospital, Heidelberg, Germany.
  • Starrach L; HCR (Heidelberg Center for Heart Rhythm Disorders), Medical University Hospital, Heidelberg, Germany.
  • Rahm AK; DZHK (German Center for Cardiovascular Research), Partner Site Heidelberg/Mannheim, Heidelberg, Germany.
  • Thomas D; Department of Cardiology, Medical University Hospital, Heidelberg, Germany.
  • Frey N; HCR (Heidelberg Center for Heart Rhythm Disorders), Medical University Hospital, Heidelberg, Germany.
  • Lugenbiel P; DZHK (German Center for Cardiovascular Research), Partner Site Heidelberg/Mannheim, Heidelberg, Germany.
Front Cardiovasc Med ; 10: 1208250, 2023.
Article em En | MEDLINE | ID: mdl-38034374
ABSTRACT

Background:

The novel multielectrode radiofrequency (RF) balloon catheter (HELIOSTAR™, Biosense Webster) is a new technology for pulmonary vein isolation (PVI) in atrial fibrillation (AF), combining RF-ablation and 3D-mapping visualization with the concept of a "single-shot"-ablation device. This study evaluates the operator learning curve und procedural outcome during implementation of the multielectrode RF-balloon at a high-volume center.

Methods:

The first 40 patients undergoing PVI by multielectrode RF-balloon catheter at Heidelberg University Hospital were included in this prospective study. Procedural outcome was analyzed over the course of increasing experience with the device.

Results:

157/157 pulmonary veins (PVs) were successfully isolated with the RF-balloon catheter, in 73.2% by a single RF-application. Median time to isolation (TTI) was 11.0 s (Q1 = 8.0 s; Q3 = 13.8 s). Median procedure time was 62.5 min (Q1 = 50.0 min; Q3 = 70.5 min). LA-dwell time was 28.5 min (Q1 = 23.3 min; Q3 = 36.5 min). Median fluoroscopy duration was 11.6 min (Q1 = 10.1 min; Q3 = 13.7 min). No serious procedure-related complications were observed, apart from one case of unclear, post-procedural acute-on-chronic kidney injury. With increasing operator experience, an additional reduction in procedure duration was observed.

Conclusion:

Rapid implementation of a "single shot"-ablation device combining RF-ablation and 3D-mapping can be achieved with high acute procedural efficacy and safety at a high-volume center. Previous experience with "single-shot" ablation devices may be advantageous for time-efficient introduction of the novel RF-balloon catheter into clinical practice. Clinical Trial Registration ClinicalTrials.gov; Identifier NCT0560361.