Your browser doesn't support javascript.
loading
Defining the Learning Period of a Novel Imageless Navigation System for Posterior Approach Total Hip Arthroplasty: Analysis of Surgical Time and Accuracy.
Ong, Christian B; Ong, Justin M; Grubel, Jacqueline; Chiu, Yu-Fen; Premkumar, Ajay; Lee, Gwo-Chin; Della Valle, Alejandro Gonzalez.
Afiliação
  • Ong CB; The Department of Orthopaedic Surgery, Hospital for Special Surgery, 535 East 70th Street, New York, NY 10021 USA.
  • Ong JM; The Department of Orthopaedic Surgery, Hospital for Special Surgery, 535 East 70th Street, New York, NY 10021 USA.
  • Grubel J; The Department of Orthopaedic Surgery, Hospital for Special Surgery, 535 East 70th Street, New York, NY 10021 USA.
  • Chiu YF; The Department of Biostatistics, Hospital for Special Surgery, New York, NY USA.
  • Premkumar A; The Department of Orthopaedic Surgery, Hospital for Special Surgery, 535 East 70th Street, New York, NY 10021 USA.
  • Lee GC; The Department of Orthopaedic Surgery, Hospital for Special Surgery, 535 East 70th Street, New York, NY 10021 USA.
  • Della Valle AG; The Department of Orthopaedic Surgery, Hospital for Special Surgery, 535 East 70th Street, New York, NY 10021 USA.
Indian J Orthop ; 58(2): 121-126, 2024 Feb.
Article em En | MEDLINE | ID: mdl-38312909
ABSTRACT

Introduction:

The use of imageless navigation in total hip arthroplasty (THA) is frequently associated with prolonged surgical times, predominantly during the learning period. The purpose of the present study was to characterize the learning period of a novel imageless navigation system, specifically as it related to surgical time and acetabular navigation accuracy. Materials and

Methods:

This was a retrospective observational study of a consecutive group of 158 patients who underwent primary unilateral THA for osteoarthritis by a team headed by a single surgeon. All procedures used an imageless navigation system to measure acetabular cup inclination and anteversion angles, referencing a generic sagittal and frontal plane. Navigation accuracy was determined by assessing differences between intraoperative inclination and anteversion values and those obtained from standardized 6-week follow-up radiographs. Operative time and navigation accuracy were assessed by plotting moving averages of 7 consecutive cases. The learning period was defined using Mann-Kendall trend analyses, student t-tests and nonlinear regression modeling based on surgical time and navigation accuracy. Alpha error was 0.05.

Results:

The average surgical time was 67.3 min (SD9.2) (range 45-95). The average navigation accuracy for inclination was 0.01° (SD4.2) (range - 10 to 10), and that for anteversion was - 4.9° (SD3.8) (range - 14 to 5). Average surgical time and navigation accuracy were similar between the first and final cases in the series with no learning period detected.

Conclusions:

There was no discernible learning period effect on surgical time or system measurement accuracy during the early phases of adoption for this imageless navigation system.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Observational_studies Idioma: En Revista: Indian J Orthop Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Observational_studies Idioma: En Revista: Indian J Orthop Ano de publicação: 2024 Tipo de documento: Article