A multi-centre stereotactic radiosurgery planning study of multiple brain metastases using isocentric linear accelerators with 5 and 2.5 mm width multi-leaf collimators, CyberKnife and Gamma Knife.
Hanvey, Scott; Hackett, Philippa; Winch, Lucy; Lim, Elizabeth; Laney, Robin; Welsh, Liam.
Afiliação
  • Hanvey S; University Hospitals Plymouth NHS Trust, Plymouth, PL6 8DH, United Kingdom.
  • Hackett P; The Royal Marsden, London, SW3 6JJ, United Kingdom.
  • Winch L; University Hospitals Bristol NHS Foundation Trust, Bristol, BS2 8ED, United Kingdom.
  • Lim E; University Hospitals Plymouth NHS Trust, Plymouth, PL6 8DH, United Kingdom.
  • Laney R; University of Plymouth, Plymouth, PL4 8AA, United Kingdom.
  • Welsh L; University Hospitals Plymouth NHS Trust, Plymouth, PL6 8DH, United Kingdom.
BJR Open ; 6(1): tzae003, 2024 Jan.
Article em En | MEDLINE | ID: mdl-38371494
ABSTRACT

Objectives:

This study compared plans of high definition (HD), 2.5 mm width multi-leaf collimator (MLC), to standard, 5 mm width, isocentric linear accelerator (linacs), CyberKnife (CK), and Gamma Knife (GK) for stereotactic radiosurgery (SRS) techniques on multiple brain metastases.

Methods:

Eleven patients undergoing SRS for multiple brain metastases were chosen. Targets and organs at risk (OARs) were delineated and optimized SRS plans were generated and compared.

Results:

The linacs delivered similar conformity index (CI) values, but the gradient index (GI) for HD MLCs was significantly lower (P-value <.001). Half the OARs received significantly lower dose using HD MLCs. CK delivered a significantly lower CI than HD MLC linac (P-value <.001), but a significantly higher GI (P-value <.001). CI was significantly improved with the HD MLC linac compared to GK (P-value = 4.591 × 10-3), however, GK delivered a significantly lower GI (P-value <.001). OAR dose sparing was similar for the HD MLC TL, CK, and GK.

Conclusions:

Comparing linacs for SRS, the preferred choice is HD MLCs. Similar results were achieved with the HD MLC linac, CK, or GK, with each delivering significant improvements in different aspects of plan quality. Advances in knowledge This article is the first to compare HD and standard width MLC linac plans using a combination of single isocentre volumetric modulated arc therapy and multi-isocentric dynamic conformal arc plans as required, which is a more clinically relevant assessment. Furthermore, it compares these plans with CK and GK, assessing the relative merits of each technique.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: BJR Open Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Reino Unido