Determination of sensed and paced atrial-ventricular delay in cardiac resynchronization therapy.
Bank, Alan J; Brown, Christopher D; Burns, Kevin V; Johnson, Katie M.
Afiliação
  • Bank AJ; Research Department, Minneapolis Heart Institute East at United Hospital, St. Paul, Minnesota, USA.
  • Brown CD; Heart Rhythm Science Center, Minneapolis Heart Institute Foundation, Minneapolis, Minnesota, USA.
  • Burns KV; Research Department, Minneapolis Heart Institute East at United Hospital, St. Paul, Minnesota, USA.
  • Johnson KM; Research Department, Minneapolis Heart Institute East at United Hospital, St. Paul, Minnesota, USA.
Pacing Clin Electrophysiol ; 47(4): 533-541, 2024 04.
Article em En | MEDLINE | ID: mdl-38477034
ABSTRACT

BACKGROUND:

Optimization of atrial-ventricular delay (AVD) during atrial sensing (SAVD) and pacing (PAVD) provides the most effective cardiac resynchronization therapy (CRT). We demonstrate a novel electrocardiographic methodology for quantifying electrical synchrony and optimizing SAVD/PAVD.

METHODS:

We studied 40 CRT patients with LV activation delay. Atrial-sensed to RV-sensed (As-RVs) and atrial-paced to RV-sensed (Ap-RVs) intervals were measured from intracardiac electrograms (IEGM). LV-only pacing was performed over a range of SAVD/PAVD settings. Electrical dyssynchrony (cardiac resynchronization index; CRI) was measured at each setting using a multilead ECG system placed over the anterior and posterior torso. Biventricular pacing, which included multiple interventricular delays, was also conducted in a subset of 10 patients.

RESULTS:

When paced LV-only, peak CRI was similar (93 ± 5% vs. 92 ± 5%) during atrial sensing or pacing but optimal PAVD was 61 ± 31 ms greater than optimal SAVD. The difference between As-RVs and Ap-RVs intervals on IEGMs (62 ± 31 ms) was nearly identical. The slope of the correlation line (0.98) and the correlation coefficient r (0.99) comparing the 2 methods of assessing SAVD-PAVD offset were nearly 1 and the y-intercept (0.63 ms) was near 0. During simultaneous biventricular (BiV) pacing at short AVD, SAVD and PAVD programming did not affect CRI, but CRI was significantly (p < .05) lower during atrial sensing at long AVD.

CONCLUSIONS:

A novel methodology for measuring electrical dyssynchrony was used to determine electrically optimal SAVD/PAVD during LV-only pacing. When BiV pacing, shorter AVDs produce better electrical synchrony.