Contouring aid tools in radiotherapy. Smoothing: the false friend.
Marruecos Querol, Jordi; Jurado-Bruggeman, Diego; Lopez-Vidal, Anna; Mesía Nin, Ricard; Rubió-Casadevall, Jordi; Buxó, Maria; Eraso Urien, Aranzazu.
Afiliação
  • Marruecos Querol J; Radiation Oncology Department, Catalan Institute of Oncology, Girona, Spain. jmarruecos@iconcologia.net.
  • Jurado-Bruggeman D; Research Group in Radiation Oncology and Medical Physics of Girona, Girona Biomedical Research Institute (IDIBGI), Girona, Spain. jmarruecos@iconcologia.net.
  • Lopez-Vidal A; Department of Radiation Oncology, ICO, Girona, Spain. jmarruecos@iconcologia.net.
  • Mesía Nin R; Research Group in Radiation Oncology and Medical Physics of Girona, Girona Biomedical Research Institute (IDIBGI), Girona, Spain.
  • Rubió-Casadevall J; Medical Physics and Radiation Protection Department, Catalan Institute of Oncology, Girona, Spain.
  • Buxó M; Medical Oncology Department, Catalan Institute of Oncology, Girona, Spain.
  • Eraso Urien A; Medical Oncology Department, Catalan Institute of Oncology, B-ARGO Group, IGTP, Badalona, Spain.
Clin Transl Oncol ; 26(8): 1956-1967, 2024 Aug.
Article em En | MEDLINE | ID: mdl-38493446
ABSTRACT

OBJECTIVE:

Contouring accuracy is critical in modern radiotherapy. Several tools are available to assist clinicians in this task. This study aims to evaluate the performance of the smoothing tool in the ARIA system to obtain more consistent volumes.

METHODS:

Eleven different geometric shapes were delineated in ARIA v15.6 (Sphere, Cube, Square Prism, Six-Pointed Star Prism, Arrow Prism, And Cylinder and the respective volumes at 45° of axis deviation (_45)) in 1, 3, 5, 7, and 10 cm side or diameter each. Post-processing drawing tools to smooth those first-generated volumes were applied in different options (2D-ALL vs 3D) and grades (1, 3, 5, 10, 15, and 20). These volumetric transformations were analyzed by comparing different parameters volume changes, center of mass, and DICE similarity coefficient index. Then we studied how smoothing affected two different volumes in a head and neck cancer patient a single rounded node and the volume delineating cervical nodal areas.

RESULTS:

No changes in data were found between 2D-ALL or 3D smoothing. Minimum deviations were found (range from 0 to 0.45 cm) in the center of mass. Volumes and the DICE index decreased as the degree of smoothing increased. Some discrepancies were found, especially in figures with cleft and spikes that behave differently. In the clinical case, smoothing should be applied only once throughout the target delineation process, preferably in the largest volume (PTV) to minimize errors.

CONCLUSION:

Smoothing is a good tool to reduce artifacts due to the manual delineation of radiotherapy volumes. The resulting volumes must be always carefully reviewed.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Neoplasias de Cabeça e Pescoço Limite: Humans Idioma: En Revista: Clin Transl Oncol Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Espanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Neoplasias de Cabeça e Pescoço Limite: Humans Idioma: En Revista: Clin Transl Oncol Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Espanha