Minimizing electric fields and increasing peripheral nerve stimulation thresholds using a body gradient array coil.
Babaloo, Reza; Atalar, Ergin.
Afiliação
  • Babaloo R; National Magnetic Resonance Research Center (UMRAM), Bilkent University, Ankara, Turkey.
  • Atalar E; Department of Electrical and Electronics Engineering, Bilkent University, Ankara, Turkey.
Magn Reson Med ; 92(3): 1290-1305, 2024 Sep.
Article em En | MEDLINE | ID: mdl-38624032
ABSTRACT

PURPOSE:

To demonstrate the performance of gradient array coils in minimizing switched-gradient-induced electric fields (E-fields) and improving peripheral nerve stimulation (PNS) thresholds while generating gradient fields with adjustable linearity across customizable regions of linearity (ROLs).

METHODS:

A body gradient array coil is used to reduce the induced E-fields on the surface of a body model by modulating applied currents. This is achieved by performing an optimization problem with the peak E-field as the objective function and current amplitudes as unknown variables. Coil dimensions and winding patterns are fixed throughout the optimization, whereas other engineering metrics remain adjustable. Various scenarios are explored by manipulating adjustable parameters.

RESULTS:

The array design consistently yields lower E-fields and higher PNS thresholds across all scenarios compared with a conventional coil. When the gradient array coil generates target gradient fields within a 44-cm-diameter spherical ROL, the maximum E-field is reduced by 10%, 18%, and 61% for the X, Y, and Z gradients, respectively. Transitioning to a smaller ROL (24 cm) and relaxing the gradient linearity error results in further E-field reductions. In oblique gradients, the array coil demonstrates the most substantial reduction of 40% in the Z-Y direction. Among the investigated scenarios, the most significant increase of 4.3-fold is observed in the PNS thresholds.

CONCLUSION:

Our study demonstrated that gradient array coils offer a promising pathway toward achieving high-performance gradient coils regarding gradient strength, slew rate, and PNS thresholds, especially in scenarios in which linear magnetic fields are required within specific target regions.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Nervos Periféricos / Desenho de Equipamento Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Turquia

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Nervos Periféricos / Desenho de Equipamento Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Turquia