Severe Hemodynamic Collapse During Humerus Stabilization with Photodynamic Implant: A Report of Two Cases.
JBJS Case Connect
; 14(2)2024 Apr 01.
Article
em En
| MEDLINE
| ID: mdl-38635765
ABSTRACT
CASE We present 2 cases of severe hemodynamic collapse during prophylactic stabilization of impending pathologic humerus fractures using a photodynamic bone stabilization device. Both events occurred when the monomer was infused under pressure into a balloon catheter. CONCLUSION:
We suspect that an increase in intramedullary pressure during balloon expansion may cause adverse systemic effects similar to fat embolism or bone cement implantation syndrome. Appropriate communication with the anesthesia team, invasive hemodynamic monitoring, and prophylactic vent hole creation may help mitigate or manage these adverse systemic effects.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Doenças Vasculares
/
Embolia Gordurosa
/
Fraturas Espontâneas
Limite:
Humans
Idioma:
En
Revista:
JBJS Case Connect
Ano de publicação:
2024
Tipo de documento:
Article