Design and Simulation of a 19-Electrode MEMS Piezoelectric Thin-Film Micro-Deformable Mirror for Ophthalmology.
Hu, Yisen; Yin, Hongbo; Li, Maoying; Bai, Tianyu; He, Liang; Hu, Zhimin; Xia, Yuanlin; Wang, Zhuqing.
Afiliação
  • Hu Y; School of Mechanical Engineering, Sichuan University, Chengdu 610065, China.
  • Yin H; Key Laboratory of Radiation Physics and Technology of Ministry of Education, Institute of Nuclear Science and Technology, Sichuan University, Chengdu 610064, China.
  • Li M; Med+X Center for Manufacturing, West China Hospital, Sichuan University, Chengdu 610041, China.
  • Bai T; Department of Ophthalmology, West China Hospital, Sichuan University, Chengdu 610041, China.
  • He L; School of Mechanical Engineering, Sichuan University, Chengdu 610065, China.
  • Hu Z; School of Mechanical Engineering, Sichuan University, Chengdu 610065, China.
  • Xia Y; Med+X Center for Manufacturing, West China Hospital, Sichuan University, Chengdu 610041, China.
  • Wang Z; School of Mechanical Engineering, Sichuan University, Chengdu 610065, China.
Micromachines (Basel) ; 15(4)2024 Apr 17.
Article em En | MEDLINE | ID: mdl-38675350
ABSTRACT
This study presents a numerical simulation-based investigation of a MEMS (micro-electromechanical systems)technology-based deformable mirror employing a piezoelectric film for fundus examination in adaptive optics. Compared to the classical equal-area electrode arrangement model, we optimize the electrode array for higher-order aberrations. The optimized model centralizes electrodes around the mirror center, which realizes low-voltage driving with high-accuracy correction. The optimized models exhibited commendable correction abilities, achieving a unidirectional displacement of 5.74 µm with a driven voltage of 15 V. The voltage-displacement relationship demonstrated high linearity at 0.99. Furthermore, the deformable mirror's influence matrix was computed, aligning with the Zernike standard surface shape of the order 1-3. To quantify aberration correction capabilities, fitting residuals for both models were calculated. The results indicate an average removal of 96.8% of aberrations to the human eye. This underscores that the optimized model outperforms the classical model in correcting high-order aberrations.