Optical diagnosis in still images of colorectal polyps: comparison between expert endoscopists and PolyDeep, a Computer-Aided Diagnosis system.
Davila-Piñón, Pedro; Nogueira-Rodríguez, Alba; Díez-Martín, Astrid Irene; Codesido, Laura; Herrero, Jesús; Puga, Manuel; Rivas, Laura; Sánchez, Eloy; Fdez-Riverola, Florentino; Glez-Peña, Daniel; Reboiro-Jato, Miguel; López-Fernández, Hugo; Cubiella, Joaquín.
Afiliação
  • Davila-Piñón P; Research Group in Gastrointestinal Oncology Ourense, Hospital Universitario de Ourense, Ourense, Spain.
  • Nogueira-Rodríguez A; Fundación Pública Galega de Investigación Biomédica Galicia Sur, Complexo Hospitalario Universitario de Ourense, Sergas, Ourense, Spain.
  • Díez-Martín AI; Department of Computer Science, Escuela Superior de Ingenieria Informática (ESEI), CINBIO, University of Vigo, Ourense, Spain.
  • Codesido L; Next Generation Computer Systems Group (SING) Research Group, Galicia Sur Health Research Institute (IIS Galicia Sur), Ourense, Spain.
  • Herrero J; Research Group in Gastrointestinal Oncology Ourense, Hospital Universitario de Ourense, Ourense, Spain.
  • Puga M; Fundación Pública Galega de Investigación Biomédica Galicia Sur, Complexo Hospitalario Universitario de Ourense, Sergas, Ourense, Spain.
  • Rivas L; Research Group in Gastrointestinal Oncology Ourense, Hospital Universitario de Ourense, Ourense, Spain.
  • Sánchez E; Fundación Pública Galega de Investigación Biomédica Galicia Sur, Complexo Hospitalario Universitario de Ourense, Sergas, Ourense, Spain.
  • Fdez-Riverola F; Research Group in Gastrointestinal Oncology Ourense, Hospital Universitario de Ourense, Ourense, Spain.
  • Glez-Peña D; Department of Gastroenterology, Hospital Universitario de Ourense, Ourense, Spain.
  • Reboiro-Jato M; Department of Gastroenterology, Hospital Universitario de Ourense, Centro de Investigación Biomédica en Red de Enfermedades Hepáticas y Digestivas (CIBEREHD), Ourense, Spain.
  • López-Fernández H; Research Group in Gastrointestinal Oncology Ourense, Hospital Universitario de Ourense, Ourense, Spain.
  • Cubiella J; Department of Gastroenterology, Hospital Universitario de Ourense, Ourense, Spain.
Front Oncol ; 14: 1393815, 2024.
Article em En | MEDLINE | ID: mdl-38846970
ABSTRACT

Background:

PolyDeep is a computer-aided detection and classification (CADe/x) system trained to detect and classify polyps. During colonoscopy, CADe/x systems help endoscopists to predict the histology of colonic lesions.

Objective:

To compare the diagnostic performance of PolyDeep and expert endoscopists for the optical diagnosis of colorectal polyps on still images.

Methods:

PolyDeep Image Classification (PIC) is an in vitro diagnostic test study. The PIC database contains NBI images of 491 colorectal polyps with histological diagnosis. We evaluated the diagnostic performance of PolyDeep and four expert endoscopists for neoplasia (adenoma, sessile serrated lesion, traditional serrated adenoma) and adenoma characterization and compared them with the McNemar test. Receiver operating characteristic curves were constructed to assess the overall discriminatory ability, comparing the area under the curve of endoscopists and PolyDeep with the chi- square homogeneity areas test.

Results:

The diagnostic performance of the endoscopists and PolyDeep in the characterization of neoplasia is similar in terms of sensitivity (PolyDeep 89.05%; E1 91.23%, p=0.5; E2 96.11%, p<0.001; E3 86.65%, p=0.3; E4 91.26% p=0.3) and specificity (PolyDeep 35.53%; E1 33.80%, p=0.8; E2 34.72%, p=1; E3 39.24%, p=0.8; E4 46.84%, p=0.2). The overall discriminative ability also showed no statistically significant differences (PolyDeep 0.623; E1 0.625, p=0.8; E2 0.654, p=0.2; E3 0.629, p=0.9; E4 0.690, p=0.09). In the optical diagnosis of adenomatous polyps, we found that PolyDeep had a significantly higher sensitivity and a significantly lower specificity. The overall discriminative ability of adenomatous lesions by expert endoscopists is significantly higher than PolyDeep (PolyDeep 0.582; E1 0.685, p < 0.001; E2 0.677, p < 0.0001; E3 0.658, p < 0.01; E4 0.694, p < 0.0001).

Conclusion:

PolyDeep and endoscopists have similar diagnostic performance in the optical diagnosis of neoplastic lesions. However, endoscopists have a better global discriminatory ability than PolyDeep in the optical diagnosis of adenomatous polyps.