Your browser doesn't support javascript.
loading
Transitioning from 0.5 to 0.9 mT: Protecting against inadvertent activation of magnet mode in active implants.
Steckner, Michael C; Grainger, David; Charles-Edwards, Geoff.
Afiliação
  • Steckner MC; MKS Consulting, Ohio, Beachwood, USA.
  • Grainger D; IEC SC62B/MT40 IEC 60601-2-33, Geneva, Switzerland.
  • Charles-Edwards G; ISO/TC 150/SC 6/JWG 2 (IEC/SC62B), Geneva, Switzerland.
Magn Reson Med ; 92(5): 2237-2245, 2024 Nov.
Article em En | MEDLINE | ID: mdl-38968006
ABSTRACT
The "5 gauss line" is a phrase that is likely to be familiar to everyone working with MRI, but what is its significance, how was it defined, and what changes are currently in progress? This review explores the history of 5 gauss (0.5 mT) as a threshold for protecting against inadvertently putting cardiac pacemakers, implantable cardioverter defibrillators, and other active implantable medical devices into a "magnet mode." Additionally, it describes the background to the recent change of this threshold to 9 gauss (0.9 mT) in the International Standard IEC 60601-2-33 edition 4.0 that defines basic safety requirements for MRI. Practical implications of this change and some ongoing and emerging issues are also discussed.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética Limite: Humans Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos