Your browser doesn't support javascript.
loading
Contrast enhancement boost improves the image quality of CT angiography derived from 80-kVp cerebral CT perfusion data.
Yang, Lin; Zhang, Haiwei; Sheng, Jiexin; Wang, Meng; Liu, Yaliang; Xu, Min; Yang, Xiao; Wang, Bo; He, Xiaolong; Gao, Lei; Zheng, Chao.
Afiliação
  • Yang L; Department of Radiology, Hanzhong Central Hospital, Hanzhong City, Shannxi Province, China.
  • Zhang H; Department of General Medicine, Hanzhong Central Hospital, Hanzhong, China.
  • Sheng J; Department of Radiology, Hanzhong Central Hospital, Hanzhong City, Shannxi Province, China.
  • Wang M; Department of Radiology, Hanzhong Central Hospital, Hanzhong City, Shannxi Province, China.
  • Liu Y; Department of Radiology, Hanzhong Central Hospital, Hanzhong City, Shannxi Province, China.
  • Xu M; Canon Medical Systems (China), Beijing, China.
  • Yang X; Canon Medical Systems (China), Beijing, China.
  • Wang B; Department of Radiology, Hanzhong Central Hospital, Hanzhong City, Shannxi Province, China.
  • He X; Department of Radiology, Hanzhong Central Hospital, Hanzhong City, Shannxi Province, China.
  • Gao L; Department of Nneurology, Hanzhong Central Hospital, Hanzhong, China.
  • Zheng C; Department of Radiology, Hanzhong Central Hospital, Hanzhong City, Shannxi Province, China. lanlan360520@163.com.
BMC Med Imaging ; 24(1): 193, 2024 Jul 30.
Article em En | MEDLINE | ID: mdl-39080580
ABSTRACT
RATIONALE AND

OBJECTIVE:

To investigate the impact of the contrast enhancement boost (CE-boost) technique on the image quality of CT angiography (CTA) derived from 80-kVp cerebral CT perfusion (CTP) data, and to compare it with conventional CTApeak as well as other currently employed methods for enhancing CTA images, such as CTAtMIP and CTAtAve extracted from CTP. MATERIALS AND

METHODS:

The data of forty-seven patients who underwent CTP at 80 kVp were retrospectively collected. Four sets of images CTApeak, CTAtMIP, CTAtAve, and CE-boost images. The CTApeak image represents the arterial phase at its peak value, captured as a single time point. CTAtMIP and CTAtAve are 4D CTA images that provide maximum density projection and average images from the three most prominent time points. CE-boost is a postprocessing technique used to enhance contrast in the arterial phase at its peak value. We compared the average CT value, standard deviation (SD), signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) of the internal carotid artery (ICA) and basilar artery (BA) among the four groups. Image quality was evaluated using a 5-point scale.

RESULTS:

The CE-boost demonstrated and CNR in the ICA and BA (all p < 0.001). Compared with the other three CTA reconstructed images, the CE-boost images had the best subjective image quality, with the highest scores of 4.77 ± 0.43 and 4.87 ± 0.34 for each reader (all p < 0.001).

CONCLUSION:

Compared with other currently used techniques,CE-boost enhances the image quality of CTA derived from 80-kVp CTP data, leading to improved visualization of intracranial arteries.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Meios de Contraste / Razão Sinal-Ruído / Angiografia por Tomografia Computadorizada Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: BMC Med Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Meios de Contraste / Razão Sinal-Ruído / Angiografia por Tomografia Computadorizada Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: BMC Med Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China