Biodesign: Engineering an aortic endograft explantation tool.
Hatami, Solyman; Maturi, Vamsi; Mathew, Alwin; Lu, Shannon; Haddad, Paul; Sheikh, Daanish; Rahimi, Maham.
Afiliação
  • Hatami S; Texas A&M School of Engineering Medicine, Houston, TX.
  • Maturi V; Texas A&M School of Engineering Medicine, Houston, TX.
  • Mathew A; Texas A&M School of Engineering Medicine, Houston, TX.
  • Lu S; Texas A&M School of Engineering Medicine, Houston, TX.
  • Haddad P; Department of Cardiovascular Surgery, Houston Methodist, Houston, TX.
  • Sheikh D; UTHSCSA Long School of Medicine, San Antonio, TX.
  • Rahimi M; Department of Cardiovascular Surgery, Houston Methodist, Houston, TX.
J Vasc Surg Cases Innov Tech ; 10(6): 101599, 2024 Dec.
Article em En | MEDLINE | ID: mdl-39351211
ABSTRACT
Endovascular aortic repair (EVAR) graft failure can be as high as 16% to 30% owing to endoleak, graft migration, or infection, often necessitating explantation, leading to potential morbidity (31%) and mortality (6.3%). Graft prongs frequently tear through the endothelium during explantation, leading to endothelial damage and subsequent fatal bleeding. The current standard of care involves different suboptimal techniques such as the syringe technique in which a cylinder is improvised by cutting a syringe in half and pushed over the graft hooks in a rotating motion, until covered for manual explantation. Because there is no commercially available product to address this shortcoming in graft explantation, we engage in the biodesign process to produce a functional explantation device. We designed and prototyped multiple potential solutions to remove EVAR endografts safely. Silicone tubing with EVAR endografts deployed in the lumen were used to simulate a grafted aorta and test each prototype. Prototypes were compared in their ability to meet design criteria including decrease in graft diameter, prevention of arterial dissection, ease of use, and decrease in procedure time. After determining the single best prototype, surgeon feedback was elicited to iteratively improve the original design. The most effective design uses a tapered lumenal geometry that decreases the EVAR graft diameter and uses stainless steel beads to prevent shear stress to the simulated aorta. A distal grip allows for easy single hand manipulation of the device, while a latching mechanism allows for smooth placement and removal over the endograft. After rigorous prototyping, our device proved feasible and effective for safe EVAR explantation, allowing this procedure to be performed safely.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: J Vasc Surg Cases Innov Tech Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: J Vasc Surg Cases Innov Tech Ano de publicação: 2024 Tipo de documento: Article