Your browser doesn't support javascript.
loading
Use of capillary optics as a beam intensifier for a Compton x-ray source.
Tompkins, P A; Abreu, C C; Carroll, F E; Xiao, Q F; MacDonald, C A.
Afiliação
  • Tompkins PA; Department of Physics and Astronomy, Vanderbilt University, Nashville, Tennessee.
Med Phys ; 21(11): 1777-84, 1994 Nov.
Article em En | MEDLINE | ID: mdl-7891640
ABSTRACT
The use of Kumakhov capillary optics will significantly enhance the performance of near-monochromatic, Compton backscattered x-ray programs. The Vanderbilt University Medical Free-Electron Laser Center is developing the capability to create these tunable x rays for medical imaging. The present transport has only reflection optics, and the beam is quite large in diameter at the laboratory. Low loss collimation of this beam would allow higher x-ray intensities after transport. This article describes experimental and computer simulation results which predict the expected performance for a multifiber Kumakhov collimator for use in the x-ray beam transport. Estimates from our research are that a multifiber optic formed of individual polycapillary fibers could be used to capture the full 7 mrad of the Vanderbilt x-ray beam and collimate it to a 1-2 mrad divergence with approximately 40%-50% transmission efficiency. This optic should increase the x-ray intensity at the laboratory level by a factor of > or = 5 by decreasing the beam divergence and subsequent spot size. Additionally, analysis of monolithic optics of fused multicapillary fibers predicts an increase in the intensity of the x rays at the laboratory by a factor of 55. These optics can have tapered channels that greatly decrease their exit divergence. This will greatly enhance the capabilities of this unique x-ray source. This article reports the initial results from a collaboration between Vanderbilt, The Center for X-Ray Optics at University at Albany, SUNY, and X-Ray Optical Systems in Albany, NY.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ecrans Intensificadores para Raios X / Tecnologia de Fibra Óptica Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 1994 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ecrans Intensificadores para Raios X / Tecnologia de Fibra Óptica Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 1994 Tipo de documento: Article