Anaesthesist ; 62(3): 197-200, 2013 Mar.
Artículo en Alemán | MEDLINE | ID: mdl-23475174

RESUMEN

This article presents the case of a patient with massive postoperative hypercapnia during mechanical ventilation in the intensive care unit (ICU). With normal tidal volumes and clearly visible chest movements, adequate findings with regard to auscultation, oxygenation and correct respirator settings, no cause for the increasing hypercapnia was initially found; however, replacement of the respirator led to a return to normal carbon dioxide levels. When checking the replaced respirator a service technician found the cause of the respirator failure: the internal tube of the co-axial ventilation system was faulty leading to an increased dead space and rebreathing of carbon dioxide.


Asunto(s)
Anestesia por Circuito Cerrado/instrumentación , Hipercapnia/etiología , Hipercapnia/terapia , Complicaciones Posoperatorias/etiología , Complicaciones Posoperatorias/terapia , Respiración Artificial/instrumentación , Ventiladores Mecánicos , Anciano de 80 o más Años , Anestesia por Circuito Cerrado/efectos adversos , Anestesia General , Análisis de los Gases de la Sangre , Dióxido de Carbono/sangre , Cuidados Críticos , Falla de Equipo , Volumen Espiratorio Forzado , Humanos , Masculino , Terapia por Inhalación de Oxígeno , Enfermedad Pulmonar Obstructiva Crónica/complicaciones , Radiografía Torácica , Espacio Muerto Respiratorio , Volumen de Ventilación Pulmonar