Nanotechnology ; 29(50): 505705, 2018 Dec 14.
Artículo en Inglés | MEDLINE | ID: mdl-30251963

RESUMEN

Nanoparticles or similar, nanoscale objects such as proteins or biological fibrils usually have to be deposited from aqueous suspension onto a solid support surface for further characterization by atomic force microscopy (AFM) and related methods such as Kelvin-probe force microscopy (KFM). Here we show, on the examples of functionalized nanoparticles and collagen fibrils, that water desorption after sample preparation affects their electrostatic potential determined by KFM in a predictable manner. We explain this effect with a simple, analytical model based on the capacitance of the partially dielectric-filled tip-sample system. We also propose practical measures to avoid false interpretation of electrical AFM-based experiments. As the phenomenon is very generic it may have significant implications in the application of AFM to nanoparticles and other nanostructures including biological ones.