1.
Talanta ; 234: 122586, 2021 Nov 01.
Artículo en Inglés | MEDLINE | ID: mdl-34364415

RESUMEN

This paper reports the development of a method based on the extraction induced by emulsion breaking (EIEB) for the determination of Cr, Cu and Mn in asphalt by high-resolution continuum source atomic absorption spectrometry. In optimized conditions, the extraction efficiency ranged from 88.6 for Cu to 104.5% for Mn. Measurements were carried out at the primary lines of the three analytes. The limits of detection were 0.02 µg g-1 for Cr, 0.01 µg g-1 for Cu and 7.11 ng g-1 for Mn. The precision, expressed as the relative standard deviation, ranged from 1.0 for Cu to 9.1% for Cr. The performance of the proposed method was compared to previously reported sample preparation procedures, including microwave-assisted acid digestion (MW-AD), ultrasound-assisted acid extraction (USE), emulsification (E), and direct dilution in organic solvent (DD). The parameters evaluated were matrix effects, limits of detection and quantification, characteristic mass, precision, accuracy, sample stability and applicability for routine analysis. The methods most suitable for Cu determination were EIEB, MW-AD, USE and E. For Cr and Mn, the best results were obtained by EIEB, MW-AD and DD procedures.


Asunto(s)
Hidrocarburos , Emulsiones , Espectrofotometría Atómica
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA