Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Más filtros

Banco de datos
Tipo del documento
Asunto de la revista
Intervalo de año de publicación
1.
J Synchrotron Radiat ; 23(Pt 4): 909-18, 2016 07.
Artículo en Inglés | MEDLINE | ID: mdl-27359139

RESUMEN

Typical X-ray diffraction measurements are made by moving a detector to discrete positions in space and then measuring the signal at each stationary position. This step-scanning method can be time-consuming, and may induce vibrations in the measurement system when the motors are accelerated and decelerated at each position. Furthermore, diffraction information between the data points may be missed unless a fine step-scanning is used, which further increases the total measurement time. To utilize beam time efficiently, the motor acceleration and deceleration time should be minimized, and the signal-to-noise ratio should be maximized. To accomplish this, an integrated continuous-scan system was developed at the Stanford Synchrotron Radiation Lightsource (SSRL). The continuous-scan system uses an in-house integrated motor controller system and counter/timer electronics. SPEC software is used to control both the hardware and data acquisition systems. The time efficiency and repeatability of the continuous-scan system were tested using X-ray diffraction from a ZnO powder and compared with the step-scan technique. Advantages and limitations of the continuous-scan system and a demonstration of variable-velocity continuous scan are discussed.