Radiat Environ Biophys ; 56(3): 269-276, 2017 08.
Artículo en Inglés | MEDLINE | ID: mdl-28639140

RESUMEN

An LBO (Li2B4O7) walled ionization chamber was designed to monitor the epithermal neutron fluence in boron neutron capture therapy clinical irradiation. The thermal and epithermal neutron sensitivities of the device were evaluated using accelerator neutrons from the 9Be(d, n) reaction at a deuteron energy of 4 MeV (4 MeV d-Be neutrons). The response of the chamber in terms of the electric charge induced in the LBO chamber was compared with the thermal and epithermal neutron fluences measured using the gold-foil activation method. The thermal and epithermal neutron sensitivities obtained were expressed in units of pC cm2, i.e., from the chamber response divided by neutron fluence (cm-2). The measured LBO chamber sensitivities were 2.23 × 10-7 ± 0.34 × 10-7 (pC cm2) for thermal neutrons and 2.00 × 10-5 ± 0.12 × 10-5 (pC cm2) for epithermal neutrons. This shows that the LBO chamber is sufficiently sensitive to epithermal neutrons to be useful for epithermal neutron monitoring in BNCT irradiation.


Asunto(s)
Compuestos de Boro , Neutrones , Radiometría/instrumentación , Temperatura , Aceleradores de Partículas