RESUMEN
Our meticulous investigation of ePTFE graft breakage when a wire placed at the edge of an ePTFE graft was pulled, revealed that, depending on the breakage pattern, a break starts much earlier than the peak suture retention strength, which is the current international indicator for anastomotic-site break strength. Furthermore, the breakage patterns differ based on the thickness of the wire and the fiber direction of the ePTFE graft. Based on these findings, we advocate measuring the peak suture retention strength using 0.10-mm sutures and a standardized wire thickness in order to assess the anastomotic retention strength of ePTFE grafts.
Asunto(s)
Prótesis Vascular , Politetrafluoroetileno , Suturas , Microscopía Electrónica de Rastreo , Resistencia a la TracciónRESUMEN
Prosthetic valve fracture is a serious complication and may arise in patient post-valve replacement. We experienced an outlet strut fracture and leaflet escape of a Bjork-Shiley convexo-concave valve. We performed an emergency redo mitral valve replacement and successfully retrieved the fractured strut and escaped leaflet from superficial femoral artery and the abdominal aorta. The patient showed an uneventful postoperative recovery.