1.
Clin Biochem ; 42(12): 1328-30, 2009 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-19465016

RESUMEN

OBJECTIVES: To assess the performance of the Doumas bilirubin reference method. DESIGN AND METHODS: Ring trails using pooled patient specimens, a calibrator and human sera enriched with unconjugated bilirubin were analyzed in five laboratories using the Doumas bilirubin reference method. RESULTS: The coefficient of variation for the linear measurement range between laboratories ranged from 1-3%. CONCLUSIONS: The Doumas bilirubin reference method is robust and reproducible. Bilirubin results using this method may be used in the development of more accurate and reliable calibrators.


Asunto(s)
Bilirrubina/sangre , Técnicas de Laboratorio Clínico , Técnicas de Laboratorio Clínico/métodos , Técnicas de Laboratorio Clínico/normas , Estudios de Evaluación como Asunto , Humanos , Estándares de Referencia , Valores de Referencia