Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 6 de 6
Filtrar
1.
J Cardiovasc Electrophysiol ; 33(2): 154-159, 2022 02.
Artículo en Inglés | MEDLINE | ID: mdl-34953099

RESUMEN

BACKGROUND: Cardiac perforation during leadless pacemaker implantation is more likely to require intervention than perforation by a transvenous lead. This study reports the consequences of Micra pacemaker perforations and related device and operator use problems based on information the manufacturer has submitted to the Food and Drug Administration (FDA). METHODS: FDA's Manufacturer and User Facility Device Experience (MAUDE) database was searched for Micra perforations. Data extracted included deaths, major adverse clinical events (MACEs), and device and/or operator use problems. RESULTS: Between 2016 and July 2021, 563 perforations were reported within 30 days of implant and resulted in 150 deaths (27%), 499 cardiac tamponades (89%), 64 pericardial effusions (11%), and 146 patients (26%) required emergency surgery. Half of perforations were associated with 139 (25%) device problems, 78 (14%) operator use problems, and 62 (11%) combined device and operator use problems. Inadequate electrical measurements or difficult positioning were the most frequent device problems (n = 129); non-septal implants and perforation of other structures were the most frequent operator use problems (n = 69); a combined operator use and device problem resulted in 62 delivery system perforations. No device or operator use problem was identified for 282 perforations (50%), but they were associated with 78 deaths, 245 tamponades, and 57 emergency surgeries. CONCLUSION: The Micra perforations reported in MAUDE are often associated with death and major complications requiring emergency intervention. Device and use problems account for at least half of perforations. Studies are needed to identify who is at risk for a perforation and how MACE can be avoided or mitigated.


Asunto(s)
Taponamiento Cardíaco , Lesiones Cardíacas , Marcapaso Artificial , Taponamiento Cardíaco/etiología , Lesiones Cardíacas/diagnóstico por imagen , Lesiones Cardíacas/etiología , Lesiones Cardíacas/terapia , Humanos , Marcapaso Artificial/efectos adversos , Resultado del Tratamiento , Estados Unidos , United States Food and Drug Administration
2.
Am Heart J ; 218: 128-132, 2019 12.
Artículo en Inglés | MEDLINE | ID: mdl-31806088

RESUMEN

New-onset left bundle branch block (N-LBBB) after transcatheter aortic valve replacement (TAVR) is a challenging clinical dilemma. In our single-center study, 60 out of 172 patients who underwent permanent pacemaker implantation (PPM) after TAVR had N-LBBB (34.9%). At a median follow-up duration of 357 days (IQR, 178; 560 days), two patients (3.5%) were completely pacemaker-dependent, and four others (7%) were partially dependent. Twelve patients (24%) recovered conduction in their left bundle at a median follow-up duration of 5 weeks (IQR, 4; 14 weeks). Due to the lack of clinical predictors of pacemaker dependency, active surveillance is warranted and may be an alternative to permanent pacemaker implantation.


Asunto(s)
Bloqueo de Rama/terapia , Marcapaso Artificial , Complicaciones Posoperatorias/terapia , Reemplazo de la Válvula Aórtica Transcatéter/efectos adversos , Anciano , Anciano de 80 o más Años , Bloqueo de Rama/etiología , Bloqueo de Rama/mortalidad , Femenino , Estudios de Seguimiento , Humanos , Masculino , Marcapaso Artificial/estadística & datos numéricos , Complicaciones Posoperatorias/etiología , Complicaciones Posoperatorias/mortalidad , Implantación de Prótesis/estadística & datos numéricos , Recuperación de la Función , Factores de Tiempo , Reemplazo de la Válvula Aórtica Transcatéter/estadística & datos numéricos , Resultado del Tratamiento
3.
J Cardiovasc Electrophysiol ; 24(6): 658-63, 2013 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-23445517

RESUMEN

INTRODUCTION: Successful reperfusion with primary percutaneous coronary intervention (PCI) can paradoxically elicit temporary vulnerability to ventricular arrhythmia. We examined whether T-wave alternans (TWA) level is correlated with nonsustained ventricular tachycardia (NSVT) incidence in association with PCI in patients with acute ST-segment elevation myocardial infarction (STEMI). METHODS AND RESULTS: We analyzed continuous 24-hour ambulatory electrocardiograms in 48 STEMI patients during and after successful primary PCI, achieving Thrombolysis in Myocardial Infarction (TIMI) grade 3 flow. TWA was measured using modified moving average method. Maximum TWA was elevated in patients with (N = 22) compared to without (N = 26) NSVT (75.1 ± 6.3 vs 49.9 ± 3.6 µV, P < 0.005) during the 22-hour monitoring period. TWA ≥ 60µV predicted NSVT with sensitivity of 77%; specificity, 73%; positive predictive value, 71%; and negative predictive value, 79%. Area under receiver operator characteristic curve (AUC) was 0.87 for maximum TWA in predicting NSVT. By comparison, ST-segment levels did not differ in patients with versus without NSVT and were not predictive (AUC = 0.52). TWA was elevated prior to PCI and remained elevated at 30 minutes after balloon inflation despite restoration of TIMI grade 3 flow in all patients, declining by 22 hours (P < 0.05). Maximum ST-segment levels decreased from before PCI to 30 minutes after balloon inflation. TWA is regionally specific, with higher values prior to PCI in precordial lead V5 than in V1 for left coronary lesions. CONCLUSIONS: TWA may be useful in identifying individuals at heightened risk for arrhythmia in association with primary PCI and can potentially signal time-dependent changes in arrhythmia vulnerability.


Asunto(s)
Electrocardiografía , Infarto del Miocardio/cirugía , Intervención Coronaria Percutánea , Taquicardia Ventricular/etiología , Electrocardiografía Ambulatoria , Femenino , Humanos , Masculino , Persona de Mediana Edad , Complicaciones Posoperatorias , Taquicardia Ventricular/fisiopatología
4.
Heart Rhythm ; 18(7): 1132-1139, 2021 07.
Artículo en Inglés | MEDLINE | ID: mdl-33713856

RESUMEN

BACKGROUND: Leadless intracardiac pacemakers were developed to avoid the complications of transvenous pacing systems. The Medtronic Micra™ transcatheter pacemaker is one such system. We found an unexpected number of major adverse clinical events (MACE) in the Food and Drug Administration's Manufacturers and User Facility Device Experience (MAUDE) database associated with Micra implantation. OBJECTIVE: The purpose of this study was to describe these MACE and compare them to implant procedure MACE in MAUDE for Medtronic CapSureFix™ active-fixation transvenous pacing leads. METHODS: During January 2021, we queried the MAUDE database for reports of MACE for Micra pacemakers and CapSureFix leads using the simple search terms "death," "tamponade," and "perforation." Reports from 2016-2020 were included. RESULTS: The search identified 363 MACE for Micra and 960 MACE for CapSureFix leads, including 96 Micra deaths (26.4%) vs 23 CapSureFix deaths (2.4%) (P <.001); 287 Micra tamponades (79.1%) vs 225 tamponades for CapSureFix (23.4%) (P <.001); and 99 rescue thoracotomies for Micra (27.3%) vs 50 rescue thoracotomies for CapSureFix (5.2%) (P <.001). More Micra patients required cardiopulmonary resuscitation (21.8% vs 1.1%) and suffered hypotension or shock (22.0% vs 5.8%) than CapSureFix recipients (P <.001). Micra patients were more likely to survive a myocardial perforation or tear if they had surgical repair (P = .014). CONCLUSION: Micra leadless pacemaker implantation may be complicated by myocardial and vascular perforations and tears that result in cardiac tamponade and death. We estimate the incidence is low (<1%). Rescue surgery to repair perforations may be lifesaving. MACE are significantly less for implantation of CapSureFix transvenous ventricular pacing leads.


Asunto(s)
Arritmias Cardíacas/terapia , Taponamiento Cardíaco/etiología , Marcapaso Artificial/efectos adversos , Diseño de Equipo , Estudios de Seguimiento , Humanos , Estudios Retrospectivos , Factores de Riesgo , Factores de Tiempo
5.
J Interv Card Electrophysiol ; 62(3): 507-518, 2021 Dec.
Artículo en Inglés | MEDLINE | ID: mdl-33387130

RESUMEN

PURPOSE: We hypothesized that data in manufacturers' product performance reports (PPRs) can provide clinically valuable ICD and cardiac resynchronization defibrillator (CRT-D) reliability and longevity information. METHODS: Data were obtained from 2019 PPRs. Kaplan-Meier (K-M) probabilities of freedom from malfunction, normal battery depletion (NBD), and NBD + malfunction were calculated for ICD and CRT-D pulse generators (PGs) with LiMnO2 or LiSVO/CFx batteries marketed in the USA from 2010 to 2019 and compared using the log-rank test. Malfunctions (MAL) included PGs that were found outside specifications. RESULTS: Study population included 1,149,803 ICD and CRT-D PGs: Abbott (ABT; 35.1%), Biotronik (BIO; 4.6%), Boston Scientific (BSC; 23.5%), and Medtronic (MDT; 36.9%). Significant differences in reliability (p < 0.001), defined by freedom from MAL, were found between manufacturers; the majority of 6808 MAL occurred in ABT devices (n = 4045; 59.4%), followed by BSC (n = 2384; 35.0%), MDT (n = 338;5.0%), and BIO (n = 41; 0.6%). Battery failure (n = 890; 57.9%) was the most common cause of MAL compromising therapy; analysis of unique ABT battery MAL-indicated problem appeared a year prior to advisory. Significant differences (p < 0.001) in battery longevity, as defined by freedom from NBD, were found between manufacturers. Overall performance (freedom from NBD + MAL) favored BSC for CRT-D PGs and MDT and BIO for ICDs. BSC subcutaneous ICD reliability was inferior to its transvenous ICD (p < 0.001). CONCLUSION: PPRs contain valuable data that can be aggregated and analyzed to inform physicians. Differences in product reliability exist between manufacturers. Battery longevity has improved, but MAL have significantly impacted performance. PPR data may be useful for assessing product problems and new technology.


Asunto(s)
Desfibriladores Implantables , Remoción de Dispositivos , Falla de Equipo , Humanos , Reproducibilidad de los Resultados , Factores de Tiempo
6.
Heart Rhythm ; 5(6): 775-9, 2008 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-18534361

RESUMEN

BACKGROUND: The Medtronic Sprint Fidelis implantable cardioverter-defibrillator high-voltage lead is prone to fracture. The October 2007 safety advisory recommended lead impedance monitoring to aid in identifying lead fractures. OBJECTIVE: The aim of this single-center study was to examine the effectiveness of impedance monitoring for detecting Sprint Fidelis lead failures before they caused adverse clinical events such as inappropriate shocks. METHODS: Impedance and sensing information were acquired during routine clinic and CareLink follow-up and at the time of lead failure using the Patient Alert, sensing integrity counter, nonsustained episode, and electrogram features in Medtronic pulse generators. RESULTS: Between September 2004 and February 2008, 17 of 514 Sprint Fidelis leads (3.3%) followed up at our center failed between 11 and 35 months after implantation (mean 23.0 +/- 8.0 months). Fifteen of these failures (88%) were caused by pace-sense conductor fractures, and 2 (12%) were caused by high-voltage conductor defects. Twelve of 15 patients (80%) with pace-sense conductor fractures received inappropriate shocks; of these, 4 had no significant increase in lead impedance before they were shocked, 2 were shocked <3 hours after their lead impedances exceeded the 1,000 Omega audible alert threshold, and 2 patients did not hear the alarm. All pace-sense conductor failures whose sensing function could be evaluated (13 of 15) had oversensing based on stored data, and oversensing usually (11 of 13) preceded impedances changes. CONCLUSION: Impedance monitoring did not prevent inappropriate shocks in two-thirds of our patients. Thus, pace-sense conductor impedance monitoring as currently implemented does not reliably forewarn patients of a lead malfunction. Consequently, patients who have Sprint Fidelis leads remain at risk for adverse clinical events associated with pace-sense conductor fracture.


Asunto(s)
Cardiografía de Impedancia , Desfibriladores Implantables/efectos adversos , Cardiografía de Impedancia/instrumentación , Electrodos Implantados/efectos adversos , Falla de Equipo , Análisis de Falla de Equipo , Seguridad de Equipos , Femenino , Humanos , Masculino , Estudios Retrospectivos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA