1.
J Appl Clin Med Phys ; 25(8): e14400, 2024 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-38831639

RESUMEN

BACKGROUND: Quality assurance (QA) for ultra-high dose rate (UHDR) irradiation is a crucial aspect in the emerging field of FLASH radiotherapy (FLASH-RT). This innovative treatment approach delivers radiation at UHDR, demanding careful adoption of QA protocols and procedures. A comprehensive understanding of beam properties and dosimetry consistency is vital to ensure the safe and effective delivery of FLASH-RT. PURPOSE: To develop a comprehensive pre-treatment QA program for cyclotron-based proton pencil beam scanning (PBS) FLASH-RT. Establish appropriate tolerances for QA items based on this study's outcomes and TG-224 recommendations. METHODS: A 250 MeV proton spot pattern was designed and implemented using UHDR with a 215nA nozzle beam current. The QA pattern that covers a central uniform field area, various spot spacings, spot delivery modes and scanning directions, and enabling the assessment of absolute, relative and temporal dosimetry QA parameters. A strip ionization chamber array (SICA) and an Advanced Markus chamber were utilized in conjunction with a 2 cm polyethylene slab and a range (R80) verification wedge. The data have been monitored for over 3 months. RESULTS: The relative dosimetries were compliant with TG-224. The variations of temporal dosimetry for scanning speed, spot dwell time, and spot transition time were within ± 1 mm/ms, ± 0.2 ms, and ± 0.2 ms, respectively. While the beam-to-beam absolute output on the same day reached up to 2.14%, the day-to-day variation was as high as 9.69%. High correlation between the absolute dose and dose rate fluctuations were identified. The dose rate of the central 5 × 5 cm2 field exhibited variations within 5% of the baseline value (155 Gy/s) during an experimental session. CONCLUSIONS: A comprehensive QA program for FLASH-RT was developed and effectively assesses the performance of a UHDR delivery system. Establishing tolerances to unify standards and offering direction for future advancements in the evolving FLASH-RT field.


Asunto(s)
Terapia de Protones , Garantía de la Calidad de Atención de Salud , Dosificación Radioterapéutica , Planificación de la Radioterapia Asistida por Computador , Garantía de la Calidad de Atención de Salud/normas , Terapia de Protones/métodos , Terapia de Protones/normas , Humanos , Planificación de la Radioterapia Asistida por Computador/métodos , Planificación de la Radioterapia Asistida por Computador/normas , Radioterapia de Intensidad Modulada/métodos , Radioterapia de Intensidad Modulada/normas , Radiometría/métodos , Órganos en Riesgo/efectos de la radiación , Neoplasias/radioterapia , Fantasmas de Imagen
2.
J Appl Clin Med Phys ; 25(5): e14289, 2024 May.
Artículo en Inglés | MEDLINE | ID: mdl-38319666

RESUMEN

PURPOSE: To investigate the feasibility of commissioning the 16 MeV electron FLASH Extension (FLEX) in the commercial treatment planning system (TPS) for biomedical research with cell and mouse models, and in silico treatment planning studies. METHODS: To commission the FLEX system with the electron Monte Carlo (eMC) algorithm in the commercial TPS, radiochromic film was used to measure the vendor-recommended beam data. Once the beam model was generated for the eMC algorithm, supplemental measurements were collected for validation purposes and compared against the TPS-calculated results. Additionally, the newly commissioned 16 MeV FLASH beam was compared to the corresponding 16 MeV conventional electron beam. RESULTS: The eMC algorithm effectively modeled the FLEX system. The eMC-calculated PDDs and profiles for the 16 MeV electron FLASH beam agreed with measured values within 1%, on average, for 6 × 6 cm2 and 10 × 10 cm2 applicators. Flatness and symmetry deviated by less than 1%, while FWHM and penumbra agreed within 1 mm for both eMC-calculated and measured profiles. Additionally, the small field (i.e., 2-cm diameter cutout) that was measured for validation purposes agreed with TPS-calculated results within 1%, on average, for both the PDD and profiles. The FLASH and conventional dose rate 16 MeV electron beam were in agreement in regard to energy, but the profiles for larger field sizes began to deviate (>10 × 10 cm2) due to the forward-peaked nature of the FLASH beam. For cell irradiation experiments, the measured and eMC-calculated in-plane and cross-plane absolute dose profiles agreed within 1%, on average. CONCLUSIONS: The FLEX system was successfully commissioned in the commercial TPS using the eMC algorithm, which accurately modeled the forward-peaked nature of the FLASH beam. A commissioned TPS for FLASH will be useful for pre-clinical cell and animal studies, as well as in silico FLASH treatment planning studies for future clinical implementation.


Asunto(s)
Algoritmos , Electrones , Método de Montecarlo , Dosificación Radioterapéutica , Planificación de la Radioterapia Asistida por Computador , Planificación de la Radioterapia Asistida por Computador/métodos , Ratones , Humanos , Animales , Fantasmas de Imagen , Radioterapia de Intensidad Modulada/métodos , Simulación por Computador
3.
J Appl Clin Med Phys ; 25(8): e14451, 2024 Aug.
Artículo en Inglés | MEDLINE | ID: mdl-38952057

RESUMEN

PURPOSE: This study investigated the potential of a commercially available plastic scintillator, the Exradin W2, as a real-time dosimeter for ultra-high-dose-rate (UHDR) electron beams. This work aimed to characterize this system's performance under UHDR conditions and addressed limitations inherent to other conventional dosimetry systems. METHODS AND MATERIALS: We assessed the W2's performance as a UHDR electron dosimeter using a 16 MeV UHDR electron beam from the FLASH research extension (FLEX) system. Additionally, the vendor provided a beta firmware upgrade to better handle the processing of the high signal generated in the UHDR environment. We evaluated the W2 regarding dose-per-pulse, pulse repetition rate, charge versus distance, and pulse linearity. Absorbed dose measurements were compared against those from a plane-parallel ionization chamber, optically stimulated luminescent dosimeters and radiochromic film. RESULTS: We observed that the 1 × 1 mm W2 scintillator with the MAX SD was more suitable for UHDR dosimetry compared to the 1 × 3 mm W2 scintillator, capable of matching film measurements within 2% accuracy for dose-per-pulse up to 3.6 Gy/pulse. The W2 accurately ascertained the inverse square relationship regarding charge versus virtual source distance with R2 of ∼1.00 for all channels. Pulse linearity was accurately measured with the W2, demonstrating a proportional response to the delivered pulse number. There was no discernible impact on the measured charge of the W2 when switching between the available repetition rates of the FLEX system (18-180 pulses/s), solidifying consistent beam output across pulse frequencies. CONCLUSIONS: This study tested a commercial plastic scintillator detector in a UHDR electron beam, paving the way for its potential use as a real-time, patient-specific dosimetry tool for future FLASH radiotherapy treatments. Further research is warranted to test and improve the signal processing of the W2 dosimetry system to accurately measure in UHDR environments using exceedingly high dose-per-pulse and pulse numbers.


Asunto(s)
Electrones , Plásticos , Conteo por Cintilación , Plásticos/química , Conteo por Cintilación/instrumentación , Conteo por Cintilación/métodos , Humanos , Radiometría/métodos , Radiometría/instrumentación , Dosificación Radioterapéutica , Fantasmas de Imagen , Aceleradores de Partículas/instrumentación
4.
J Appl Clin Med Phys ; 25(2): e14159, 2024 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-37735808

RESUMEN

PURPOSE: Radiotherapy delivered at ultra-high-dose-rates (≥40 Gy/s), that is, FLASH, has the potential to effectively widen the therapeutic window and considerably improve the care of cancer patients. The underlying mechanism of the FLASH effect is not well understood, and commercial systems capable of delivering such dose rates are scarce. The purpose of this study was to perform the initial acceptance and commissioning tests of an electron FLASH research product for preclinical studies. METHODS: A linear accelerator (Clinac 23EX) was modified to include a non-clinical FLASH research extension (the Clinac-FLEX system) by Varian, a Siemens Healthineers company (Palo Alto, CA) capable of delivering a 16 MeV electron beam with FLASH and conventional dose rates. The acceptance, commissioning, and dosimetric characterization of the FLEX system was performed using radiochromic film, optically stimulated luminescent dosimeters, and a plane-parallel ionization chamber. A radiation survey was conducted for which the shielding of the pre-existing vault was deemed sufficient. RESULTS: The Clinac-FLEX system is capable of delivering a 16 MeV electron FLASH beam of approximately 1 Gy/pulse at isocenter and reached a maximum dose rate >3.8 Gy/pulse near the upper accessory mount on the linac gantry. The percent depth dose curves of the 16 MeV FLASH and conventional modes for the 10 × 10 cm2 applicator agreed within 0.5 mm at a range of 50% of the maximum dose. Their respective profiles agreed well in terms of flatness but deviated for field sizes >10 × 10 cm2 . The output stability of the FLASH system exhibited a dose deviation of <1%. Preliminary cell studies showed that the FLASH dose rate (180 Gy/s) had much less impact on the cell morphology of 76N breast normal cells compared to the non-FLASH dose rate (18 Gy/s), which induced large-size cells. CONCLUSION: Our studies characterized the non-clinical Clinac-FLEX system as a viable solution to conduct FLASH research that could substantially increase access to ultra-high-dose-rate capabilities for scientists.


Asunto(s)
Electrones , Radiometría , Humanos , Dosificación Radioterapéutica , Aceleradores de Partículas , Dosímetros de Radiación
5.
Int J Mol Sci ; 25(5)2024 Feb 22.
Artículo en Inglés | MEDLINE | ID: mdl-38473799

RESUMEN

Major strides have been made in the development of FLASH radiotherapy (FLASH RT) in the last ten years, but there are still many obstacles to overcome for transfer to the clinic to become a reality. Although preclinical and first-in-human clinical evidence suggests that ultra-high dose rates (UHDRs) induce a sparing effect in normal tissue without modifying the therapeutic effect on the tumor, successful clinical translation of FLASH-RT depends on a better understanding of the biological mechanisms underpinning the sparing effect. Suitable in vitro studies are required to fully understand the radiobiological mechanisms associated with UHDRs. From a technical point of view, it is also crucial to develop optimal technologies in terms of beam irradiation parameters for producing FLASH conditions. This review provides an overview of the research progress of FLASH RT and discusses the potential challenges to be faced before its clinical application. We critically summarize the preclinical evidence and in vitro studies on DNA damage following UHDR irradiation. We also highlight the ongoing developments of technologies for delivering FLASH-compliant beams, with a focus on laser-driven plasma accelerators suitable for performing basic radiobiological research on the UHDR effects.


Asunto(s)
Daño del ADN , Motivación , Humanos , Cabeza , Plasma , Radiobiología
6.
Proc Natl Acad Sci U S A ; 116(22): 10943-10951, 2019 05 28.
Artículo en Inglés | MEDLINE | ID: mdl-31097580

RESUMEN

Here, we highlight the potential translational benefits of delivering FLASH radiotherapy using ultra-high dose rates (>100 Gy⋅s-1). Compared with conventional dose-rate (CONV; 0.07-0.1 Gy⋅s-1) modalities, we showed that FLASH did not cause radiation-induced deficits in learning and memory in mice. Moreover, 6 months after exposure, CONV caused permanent alterations in neurocognitive end points, whereas FLASH did not induce behaviors characteristic of anxiety and depression and did not impair extinction memory. Mechanistic investigations showed that increasing the oxygen tension in the brain through carbogen breathing reversed the neuroprotective effects of FLASH, while radiochemical studies confirmed that FLASH produced lower levels of the toxic reactive oxygen species hydrogen peroxide. In addition, FLASH did not induce neuroinflammation, a process described as oxidative stress-dependent, and was also associated with a marked preservation of neuronal morphology and dendritic spine density. The remarkable normal tissue sparing afforded by FLASH may someday provide heretofore unrealized opportunities for dose escalation to the tumor bed, capabilities that promise to hasten the translation of this groundbreaking irradiation modality into clinical practice.


Asunto(s)
Disfunción Cognitiva , Neuroprotección/efectos de la radiación , Dosis de Radiación , Radioterapia/métodos , Especies Reactivas de Oxígeno/metabolismo , Animales , Encéfalo/patología , Encéfalo/efectos de la radiación , Disfunción Cognitiva/etiología , Disfunción Cognitiva/prevención & control , Femenino , Inflamación , Ratones , Ratones Endogámicos C57BL , Radioterapia/efectos adversos , Especies Reactivas de Oxígeno/análisis
7.
Int J Mol Sci ; 23(6)2022 Mar 09.
Artículo en Inglés | MEDLINE | ID: mdl-35328377

RESUMEN

The impact of the exact temporal pulse structure on the potential cell and tissue sparing of ultra-high dose-rate irradiation applied in FLASH studies has gained increasing attention. A previous version of our biophysical mechanistic model (UNIVERSE: UNIfied and VERSatile bio response Engine), based on the oxygen depletion hypothesis, has been extended in this work by considering oxygen-dependent damage fixation dynamics on the sub-milliseconds scale and introducing an explicit implementation of the temporal pulse structure. The model successfully reproduces in vitro experimental data on the fast kinetics of the oxygen effect in irradiated mammalian cells. The implemented changes result in a reduction in the assumed amount of oxygen depletion. Furthermore, its increase towards conventional dose-rates is parameterized based on experimental data from the literature. A recalculation of previous benchmarks shows that the model retains its predictive power, while the assumed amount of depleted oxygen approaches measured values. The updated UNIVERSE could be used to investigate the impact of different combinations of pulse structure parameters (e.g., dose per pulse, pulse frequency, number of pulses, etc.), thereby aiding the optimization of potential clinical application and the development of suitable accelerators.


Asunto(s)
Mamíferos , Oxígeno , Animales , Cinética , Dosificación Radioterapéutica
8.
Int J Mol Sci ; 23(21)2022 Nov 03.
Artículo en Inglés | MEDLINE | ID: mdl-36362271

RESUMEN

Ultra-high dose rate (UHDR) irradiation regimes have the potential to spare normal tissue while keeping equivalent tumoricidal capacity than conventional dose rate radiotherapy (CONV-RT). This has been called the FLASH effect. In this work, we present a new simulation framework aiming to study the production of radical species in water and biological media under different irradiation patterns. The chemical stage (heterogeneous phase) is based on a nonlinear reaction-diffusion model, implemented in GPU. After the first 1 µs, no further radical diffusion is assumed, and radical evolution may be simulated over long periods of hundreds of seconds. Our approach was first validated against previous results in the literature and then employed to assess the influence of different temporal microstructures of dose deposition in the expected biological damage. The variation of the Normal Tissue Complication Probability (NTCP), assuming the model of Labarbe et al., where the integral of the peroxyl radical concentration over time (AUC-ROO) is taken as surrogate for biological damage, is presented for different intra-pulse dose rate and pulse frequency configurations, relevant in the clinical scenario. These simulations yield that overall, mean dose rate and the dose per pulse are the best predictors of biological effects at UHDR.


Asunto(s)
Dosificación Radioterapéutica , Simulación por Computador
9.
Int J Mol Sci ; 23(20)2022 Oct 11.
Artículo en Inglés | MEDLINE | ID: mdl-36292961

RESUMEN

FLASH radiotherapy, or the delivery of a dose at an ultra-high dose rate (>40 Gy/s), has recently emerged as a promising tool to enhance the therapeutic index in cancer treatment. The remarkable sparing of normal tissues and equivalent tumor control by FLASH irradiation compared to conventional dose rate irradiation­the FLASH effect­has already been demonstrated in several preclinical models and even in a first patient with T-cell cutaneous lymphoma. However, the biological mechanisms responsible for the differential effect produced by FLASH irradiation in normal and cancer cells remain to be elucidated. This is of great importance because a good understanding of the underlying radiobiological mechanisms and characterization of the specific beam parameters is required for a successful clinical translation of FLASH radiotherapy. In this review, we summarize the FLASH investigations performed so far and critically evaluate the current hypotheses explaining the FLASH effect, including oxygen depletion, the production of reactive oxygen species, and an altered immune response. We also propose a new theory that assumes an important role of mitochondria in mediating the normal tissue and tumor response to FLASH dose rates.


Asunto(s)
Neoplasias , Humanos , Dosificación Radioterapéutica , Especies Reactivas de Oxígeno , Neoplasias/genética , Neoplasias/radioterapia , Oxígeno
10.
Rep Pract Oncol Radiother ; 27(2): 344-351, 2022.
Artículo en Inglés | MEDLINE | ID: mdl-36299375

RESUMEN

FLASH radiotherapy (RT) is a technique involving the delivery of ultra-high dose rate radiation to the target. FLASH-RT has been shown to reduce radiation-induced toxicity in healthy tissues without compromising the anti-cancer effects of treatment compared to conventional radiation therapy. In the present article, we review the published data on FLASH-RT and discuss the current state of knowledge of this novel approach. We also highlight the technological constraints and complexity of FLASH-RT and describe the physics underlying this modality, particularly how technology supports energy transfer by ionising radiation (e.g., beam on/off sequence, pulse-energy load, intervals). We emphasise that current preclinical experience is mostly based on FLASH electrons and that clinical application of FLASH-RT is very limited. The incorporation of FLASH-RT into routine clinical radiotherapy will require the development of devices capable of producing FLASH photon beams.

11.
J Appl Clin Med Phys ; 22(11): 165-171, 2021 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-34609051

RESUMEN

PURPOSE: To implement and validate a beam current transformer as a passive monitoring device on a pulsed electron beam medical linear accelerator (LINAC) for ultra-high dose rate (UHDR) irradiations in the operational range of at least 3 Gy to improve dosimetric procedures currently in use for FLASH radiotherapy (FLASH-RT) studies. METHODS: Two beam current transformers (BCTs) were placed at the exit of a medical LINAC capable of UHDR irradiations. The BCTs were validated as monitoring devices by verifying beam parameters consistency between nominal values and measured values, determining the relationship between the charge measured and the absorbed dose, and checking the short- and long-term stability of the charge-absorbed dose ratio. RESULTS: The beam parameters measured by the BCTs coincide with the nominal values. The charge-dose relationship was found to be linear and independent of pulse width and frequency. Short- and long-term stabilities were measured to be within acceptable limits. CONCLUSIONS: The BCTs were implemented and validated on a pulsed electron beam medical LINAC, thus improving current dosimetric procedures and allowing for a more complete analysis of beam characteristics. BCTs were shown to be a valid method for beam monitoring for UHDR (and therefore FLASH) experiments.


Asunto(s)
Electrones , Monitoreo de Radiación , Protocolos Clínicos , Humanos , Aceleradores de Partículas , Radiometría , Dosificación Radioterapéutica
12.
J Appl Clin Med Phys ; 22(6): 50-59, 2021 Jun.
Artículo en Inglés | MEDLINE | ID: mdl-34028969

RESUMEN

PURPOSE: To assess dosimetric properties and identify required updates to commonly used protocols (including use of film and ionization chamber) pertaining to a clinical linac configured into FLASH (ultra-high dose rate) electron mode. METHODS: An 18MV photon beam of a Varian iX linac was converted to FLASH electron beam by replacing the target and the flattening filter with an electron scattering foil. The dose was prescribed by entering the MUs through the console. Fundamental beam properties, including energy, dose rate, dose reproducibility, field size, and dose rate dependence on the SAD, were examined in preparation for radiobiological experiments. Gafchromic EBT-XD film was evaluated for usability in measurements at ultra-high dose rates by comparing the measured dose to the inverse square model. Selected previously reported models of chamber efficiencies were fitted to measurements in a broad range of dose rates. RESULTS: The performance of the modified linac was found adequate for FLASH radiobiological experiments. With exception of the increase in the dose per MU on increase in the repetition rate, all fundamental beam properties proved to be in line with expectations developed with conventional linacs. The field size followed the theorem of similar triangles. The highest average dose rate (2 × 104  Gy/s) was found next to the internal monitor chamber, with the field size of FWHM = 1.5 cm. Independence of the dose readings on the dose rate (up to 2 × 104  Gy/s) was demonstrated for the EBT-XD film. A model of recombination in an ionization chamber was identified that provided good agreement with the measured chamber efficiencies for the average dose rates up to at least 2 × 103  Gy/s. CONCLUSION: Dosimetric measurements were performed to characterize a linac converted to FLASH dose rates. Gafchromic EBT-XD film and dose rate-corrected cc13 ionization chamber were demonstrated usable at FLASH dose rates.


Asunto(s)
Electrones , Aceleradores de Partículas , Dosimetría por Película , Humanos , Radiometría , Reproducibilidad de los Resultados
13.
Sensors (Basel) ; 21(17)2021 Aug 30.
Artículo en Inglés | MEDLINE | ID: mdl-34502730

RESUMEN

FLASH radiotherapy is an emerging radiotherapy technique used to spare normal tissues. It employs ultra-high dose rate radiation beams over 40 Gy/s, which is significantly higher than those of conventional radiotherapy. In this study, a fiber-optic radiation sensor (FORS) was fabricated using a plastic scintillator, an optical filter, and a plastic optical fiber to measure the ultra-high dose rate electron beams over 40 Gy/s used in FLASH radiotherapy. The radiation-induced emissions, such as Cherenkov radiation and fluorescence generated in a transmitting optical fiber, were spectrally discriminated from the light outputs of the FORS. To evaluate the linearity and dose rate dependence of the FORS, the outputs of the fiber-optic radiation sensor were measured according to distances from an electron scattering device, and the results were compared with those of an ionization chamber and radiochromic films. Finally, the percentage depth doses were obtained using the FORS as a function of depth in a water phantom. This study found that ultra-high dose rate electron beams over 40 Gy/s could be measured in real time using a FORS.


Asunto(s)
Electrones , Fibras Ópticas , Tecnología de Fibra Óptica , Fantasmas de Imagen , Radiometría , Dosificación Radioterapéutica
14.
Zhongguo Yi Liao Qi Xie Za Zhi ; 44(6): 508-512, 2020 Dec 08.
Artículo en Zh | MEDLINE | ID: mdl-33314859

RESUMEN

Flash radiotherapy is a kind of radiotherapy method using ultra-high dose rate radiation. Compared with the traditional dose rate radiotherapy, it has unique radiobiological advantages. In this paper, the principle of flash radiotherapy, the process and results of biological experiments are summarized. At the same time, the advantages and challenges of flash radiotherapy are analyzed, and the future clinical application is prospected.


Asunto(s)
Dosificación Radioterapéutica , Radioterapia/métodos , Tecnología
15.
Bull Exp Biol Med ; 166(6): 785-787, 2019 Apr.
Artículo en Inglés | MEDLINE | ID: mdl-31028581

RESUMEN

We studied the effect of low and high-dose rate photon radiation on activation of cell death by apoptosis and necrosis in malignant cell lines of lymphocytic origin Raji and Jurkat (human B and T-cell lymphomas) and normal human lymphocytes from healthy volunteers. It was shown that photon radiation with ultra-high dose rate induced significantly higher levels of "early" apoptosis and lower levels of necrosis compared to γ-radiation with dose rate used for radiation therapy.


Asunto(s)
Apoptosis/efectos de la radiación , Rayos gamma , Linfocitos/efectos de la radiación , Línea Celular Tumoral , Supervivencia Celular/efectos de la radiación , Relación Dosis-Respuesta en la Radiación , Humanos , Células Jurkat , Linfocitos/patología , Necrosis/patología , Cultivo Primario de Células
16.
Med Phys ; 51(1): 670-681, 2024 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-36939370

RESUMEN

BACKGROUND: Ultra-high dose rate (FLASH) radiation has been reported to efficiently suppress tumor growth while sparing normal tissue; however, the mechanism of the differential tissue sparing effect is still not known. Oxygen has long been known to profoundly impact radiobiological responses, and radiolytic oxygen depletion has been considered to be a possible cause or contributor to the FLASH phenomenon. PURPOSE: This work investigates the impact of tissue pO2 profiles, oxygen depletion per unit dose (g), and the oxygen concentration yielding half-maximum radiosensitization (the average of its maximum value and one) (k) in tumor and normal tissue. METHODS: We developed a model that considers the dependent relationship between oxygen depletion and change of radiosensitivity by FLASH irradiation. The model assumed that FLASH irradiation depletes intracellular oxygen more rapidly than it diffuses into the cell from the extracellular environment. Cell survival was calculated based on the linear quadratic-linear model and the radiosensitivity related parameters were adjusted in 1 Gy increments of the administered dose. The model reproduced published experimental data that were obtained with different cell lines and oxygen concentrations, and was used to analyze the impact of parameter uncertainties on the radiobiological responses. This study expands the oxygen depletion analysis of FLASH to normal human tissue and tumor based on clinically determined aggregate and individual patient pO2 profiles. RESULTS: The results show that the pO2 profile is the most essential factor that affects biological response and analyses based on the median pO2 rather than the full pO2 profile can be unreliable and misleading. Additionally, the presence of a small fraction of cells on the threshold of radiobiologic hypoxia substantially alters biological response due to FLASH oxygen depletion. We found that an increment in the k value is generally more protective of tumor than normal tissue due to a higher frequency of lower pO2 values in tumors. Variation in the g value affects the dose at which oxygen depletion impacts response, but does not alter the dose-dependent response trends, if the g value is identical in both tumor and normal tissue. CONCLUSIONS: The therapeutic efficacy of FLASH oxygen depletion is likely patient and tissue-dependent. For breast cancer, FLASH is beneficial in a minority of cases; however, in a subset of well oxygenated tumors, a therapeutic gain may be realized due to induced normal tissue hypoxia.


Asunto(s)
Neoplasias , Oxígeno , Humanos , Oxígeno/metabolismo , Tolerancia a Radiación , Neoplasias/radioterapia , Radiobiología , Hipoxia
17.
Med Phys ; 51(2): 786-798, 2024 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-38103260

RESUMEN

BACKGROUND: The first clinical trials to assess the feasibility of FLASH radiotherapy in humans have started (FAST-01, FAST-02) and more trials are foreseen. To increase comparability between trials it is important to assure treatment quality and therefore establish a standard for machine quality assurance (QA). Currently, the AAPM TG-224 report is considered as the standard on machine QA for proton therapy, however, it was not intended to be used for ultra-high dose rate (UHDR) proton beams, which have gained interest due to the observation of the FLASH effect. PURPOSE: The aim of this study is to find consensus on practical guidelines on machine QA for UHDR proton beams in transmission mode in terms of which QA is required, how they should be done, which detectors are suitable for UHDR machine QA, and what tolerance limits should be applied. METHODS: A risk assessment to determine the gaps in the current standard for machine QA was performed by an international group of medical physicists. Based on that, practical guidelines on how to perform machine QA for UHDR proton beams were proposed. RESULTS: The risk assessment clearly identified the need for additional guidance on temporal dosimetry, addressing dose rate (constancy), dose spillage, and scanning speed. In addition, several minor changes from AAPM TG-224 were identified; define required dose rate levels, the use of clinically relevant dose levels, and the use of adapted beam settings to minimize activation of detector and phantom materials or to avoid saturation effects of specific detectors. The final report was created based on discussions and consensus. CONCLUSIONS: Consensus was reached on what QA is required for UHDR scanning proton beams in transmission mode for isochronous cyclotron-based systems and how they should be performed. However, the group discussions also showed that there is a lack of high temporal resolution detectors and sufficient QA data to set appropriate limits for some of the proposed QA procedures.


Asunto(s)
Terapia de Protones , Humanos , Terapia de Protones/métodos , Ciclotrones , Protones , Consenso , Radiometría , Dosificación Radioterapéutica
18.
Front Oncol ; 14: 1340190, 2024.
Artículo en Inglés | MEDLINE | ID: mdl-38711846

RESUMEN

FLASH radiotherapy (FLASH-RT) is a novel radiotherapy approach based on the use of ultra-high dose radiation to treat malignant cells. Although tumours can be reduced or eradicated using radiotherapy, toxicities induced by radiation can compromise healthy tissues. The FLASH effect is the observation that treatment delivered at an ultra-high dose rate is able to reduce adverse toxicities present at conventional dose rates. While this novel technique may provide a turning point for clinical practice, the exact mechanisms underlying the causes or influences of the FLASH effect are not fully understood. The study presented here uses data collected from 41 experimental investigations (published before March 2024) of the FLASH effect. Searchable databases were constructed to contain the outcomes of the various experiments in addition to values of beam parameters that may have a bearing on the FLASH effect. An in-depth review of the impact of the key beam parameters on the results of the experiments was carried out. Correlations between parameter values and experimental outcomes were studied. Pulse Dose Rate had positive correlations with almost all end points, suggesting viability of FLASH-RT as a new modality of radiotherapy. The collective results of this systematic review study suggest that beam parameter qualities from both FLASH and conventional radiotherapy can be valuable for tissue sparing and effective tumour treatment.

19.
Med Phys ; 51(2): 1450-1459, 2024 Feb.
Artículo en Inglés | MEDLINE | ID: mdl-37742343

RESUMEN

BACKGROUND: The possible emergence of the FLASH effect-the sparing of normal tissue while maintaining tumor control-after irradiations at dose-rates exceeding several tens of Gy per second, has recently spurred a surge of studies attempting to characterize and rationalize the phenomenon. Investigating and reporting the dose and instantaneous dose-rate of ultra-high dose-rate (UHDR) particle radiotherapy beams is crucial for understanding and assessing the FLASH effect, towards pre-clinical application and quality assurance programs. PURPOSE: The purpose of the present work is to investigate a novel diamond-based detector system for dose and instantaneous dose-rate measurements in UHDR particle beams. METHODS: Two types of diamond detectors, a microDiamond (PTW 60019) and a diamond detector prototype specifically designed for operation in UHDR beams (flashDiamond), and two different readout electronic chains, were investigated for absorbed dose and instantaneous dose-rate measurements. The detectors were irradiated with a helium beam of 145.7 MeV/u under conventional and UHDR delivery. Dose-rate delivery records by the monitoring ionization chamber and diamond detectors were studied for single spot irradiations. Dose linearity at 5 cm depth and in-depth dose response from 2 to 16 cm were investigated for both measurement chains and both detectors in a water tank. Measurements with cylindrical and plane-parallel ionization chambers as well as Monte-Carlo simulations were performed for comparisons. RESULTS: Diamond detectors allowed for recording the temporal structure of the beam, in good agreement with the one obtained by the monitoring ionization chamber. A better time resolution of the order of few µs was observed as compared to the approximately 50 µs of the monitoring ionization chamber. Both diamonds detectors show an excellent linearity response in both delivery modalities. Dose values derived by integrating the measured instantaneous dose-rates are in very good agreement with the ones obtained by the standard electrometer readings. Bragg peak curves confirmed the consistency of the charge measurements by the two systems. CONCLUSIONS: The proposed novel dosimetric system allows for a detailed investigation of the temporal evolution of UHDR beams. As a result, reliable and accurate determinations of dose and instantaneous dose-rate are possible, both required for a comprehensive characterization of UHDR beams and relevant for FLASH effect assessment in clinical treatments.


Asunto(s)
Diamante , Helio , Diamante/química , Helio/uso terapéutico , Radiometría , Método de Montecarlo
20.
Phys Med Biol ; 69(14)2024 Jul 03.
Artículo en Inglés | MEDLINE | ID: mdl-38876112

RESUMEN

Objective. To fabricate and validate a novel focused collimator designed to spare normal tissue in a murine hemithoracic irradiation model using 250 MeV protons delivered at ultra-high dose rates (UHDRs) for preclinical FLASH radiation therapy (FLASH-RT) studies.Approach. A brass collimator was developed to shape 250 MeV UHDR protons from our Varian ProBeam. Six 13 mm apertures, of equivalent size to kV x-ray fields historically used to perform hemithorax irradiations, were precisely machined to match beam divergence, allowing concurrent hemithoracic irradiation of six mice while sparing the contralateral lung and abdominal organs. The collimated field profiles were characterized by film dosimetry, and a radiation survey of neutron activation was performed to ensure the safety of staff positioning animals.Main results. The brass collimator produced 1.2 mm penumbrae radiation fields comparable to kV x-rays used in preclinical studies. The penumbrae in the six apertures are similar, with full-width half-maxima of 13.3 mm and 13.5 mm for the central and peripheral apertures, respectively. The collimator delivered a similar dose at an average rate of 52 Gy s-1for all apertures. While neutron activation produces a high (0.2 mSv h-1) initial ambient equivalent dose rate, a parallel work-flow in which imaging and setup are performed without the collimator ensures safety to staff.Significance. Scanned protons have the greatest potential for future translation of FLASH-RT in clinical treatments due to their ability to treat deep-seated tumors with high conformality. However, the Gaussian distribution of dose in proton spots produces wider lateral penumbrae compared to other modalities. This presents a challenge in small animal pre-clinical studies, where millimeter-scale penumbrae are required to precisely target the intended volume. Offering high-throughput irradiation of mice with sharp penumbrae, our novel collimator-based platform serves as an important benchmark for enabling large-scale, cost-effective radiobiological studies of the FLASH effect in murine models.


Asunto(s)
Terapia de Protones , Animales , Ratones , Terapia de Protones/instrumentación , Terapia de Protones/métodos , Órganos en Riesgo/efectos de la radiación , Dosificación Radioterapéutica