Analysis of variable flow Doppler hemodialysis access flow measurements and comparison with ultrasound dilution.
Weitzel, W F; Rubin, J M; Leavey, S F; Swartz, R D; Dhingra, R K; Messana, J M.
Afiliación
  • Weitzel WF; Departments of Internal Medicine and Radiology, University of Michigan School of Medicine, Ann Arbor, MI, USA. weitzel@umich.edu
Am J Kidney Dis ; 38(5): 935-40, 2001 Nov.
Article en En | MEDLINE | ID: mdl-11684544
The variable flow (VF) Doppler method determines access blood flow from the pump speed-induced change in Doppler signal between the arterial and venous needles. This study evaluated 35 patients in two analyses to assess VF Doppler measurement reproducibility (54 paired measurements) and compared VF Doppler and ultrasound dilution flow measurements (24 paired measurements). VF Doppler measurement variations were 4% for access flow less than 800 mL/min (n = 17), 6% for access flow of 801 to 1,600 mL/min (n = 22), and 11% for access flow greater than 1,600 mL/min (n = 15). The mean measurement coefficient of variation was 7% for VF Doppler compared with 5% for ultrasound dilution. Correlation coefficients (r) between VF Doppler and ultrasound dilution access flow measurements were 0.79 (n = 24; P < 0.0001), 0.84 for access flow less than 2,000 mL/min (n = 20; P < 0.0001), and 0.91 for access flow less than 1,600 mL/min (n = 18, P < 0.0001). VF Doppler measurements using indicated versus measured pump flow rates correlated highly (r = 0.99; P < 0.0001). VF Doppler therefore yields reproducible access volume flow measurements that correlate with ultrasound dilution measurements. The VF Doppler method is dependent on the pump-induced change in access Doppler signal and therefore is inherently most accurate and reproducible at lower access blood flow rates. This method appears capable of determining access flow rates in the clinically useful range.
Asunto(s)
Buscar en Google
Banco de datos: MEDLINE Asunto principal: Diálisis Renal / Ultrasonografía Doppler Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: Am J Kidney Dis Año: 2001 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Banco de datos: MEDLINE Asunto principal: Diálisis Renal / Ultrasonografía Doppler Tipo de estudio: Prognostic_studies Límite: Humans Idioma: En Revista: Am J Kidney Dis Año: 2001 Tipo del documento: Article País de afiliación: Estados Unidos