Your browser doesn't support javascript.
loading
Optimization of Z-axis automatic exposure control for multidetector row CT evaluation of neck and comparison with fixed tube current technique for image quality and radiation dose.
Namasivayam, S; Kalra, M K; Pottala, K M; Waldrop, S M; Hudgins, P A.
Afiliación
  • Namasivayam S; Department of Radiology, Emory University Hospital, Atlanta, GA 30322, USA.
AJNR Am J Neuroradiol ; 27(10): 2221-5, 2006.
Article en En | MEDLINE | ID: mdl-17110699
ABSTRACT
BACKGROUND AND

PURPOSE:

Z-axis automatic exposure control (AEC) technique automatically adjusts tube current based on size of body region scanned. The purpose of the current study was to compare diagnostic acceptability, noise, and radiation exposure of multidetector row CT (MDCT) of neck performed with z-axis AEC and with fixed current. MATERIALS AND

METHODS:

Two study groups of 26 patients each underwent MDCT of neck using z-axis AEC with 8 noise index (NI), 150-440 mA, and 10 NI, 75-440 mA, respectively. A control group consisting of another 26 patients underwent MDCT of neck with fixed-current technique (300 mA). Objective noise and mean tube current-time products (mA . s) were recorded. Two radiologists evaluated images for diagnostic acceptability and subjective noise on a 5-point scale.

RESULTS:

All CT examinations of study and control groups were diagnostically acceptable, though objective noise was significantly more with z-axis AEC (shoulder NI 8, 20.6 +/- 6.2 HU; NI 10, 22.2 +/- 4.6 HU) than with fixed current (16.2 +/- 6 HU) (P = .01). There was no significant difference between AEC and fixed current in diagnostic acceptability and subjective noise (P = .22-.42). AEC resulted in significant radiation dose reduction (NI 8, 186.3 +/- 20.5 mA . s; NI 10, 158.1 +/- 21.2 mA x s), compared with fixed current (235 +/- 21.8 mA x s).

CONCLUSION:

Z-axis AEC resulted in similar subjective noise and diagnostic acceptability, with radiation dose reduction of 21% for NI of 8 and 33% for NI of 10, respectively, for MDCT evaluation of neck, compared with those of fixed current technique.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Tomografía Computarizada por Rayos X / Cuello Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: AJNR Am J Neuroradiol Año: 2006 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Tomografía Computarizada por Rayos X / Cuello Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: AJNR Am J Neuroradiol Año: 2006 Tipo del documento: Article País de afiliación: Estados Unidos