Abnormal timing delays in auditory brainstem responses evoked by bilateral cochlear implant use in children.
Gordon, Karen A; Valero, Jerome; van Hoesel, Richard; Papsin, Blake C.
Afiliación
  • Gordon KA; Cochlear Implant Laboratory, The Hospital for Sick Children and the University of Toronto, Toronto, Ontario, Canada. karen.gordon@utoronto.ca
Otol Neurotol ; 29(2): 193-8, 2008 Feb.
Article en En | MEDLINE | ID: mdl-18223446
ABSTRACT

HYPOTHESIS:

A period of unilateral implant use before bilateral implantation affects timing of brainstem processes measured by the electrically evoked auditory brainstem response (EABR).

BACKGROUND:

EABR latencies decrease with unilateral implant use potentially disrupting binaural timing cues important in auditory brainstem processing of bilateral input.

METHODS:

EABRs were evoked by electrical pulses from the left, right, and both implants simultaneously in 3 groups of children. All were initially implanted at ages younger than 3 years and had the following 1) a long delay (>2 yr [n = 16]), 2) a short delay (<1 yr [n = 15]), or 3) no delay (n = 15) between left and right ear implantation. Responses were recorded on the first day of bilateral implant use and 3 and 9 months thereafter.

RESULTS:

Relative to responses evoked in the experienced ear, the naive ear showed prolonged latency in both the EABR peaks and the binaural difference response. After 3 and 9 months of bilateral implant use, the relative prolongation decreased in the long and short delay groups, but significant differences persisted in the former. No clear differences in latencies evoked by the left versus right implant were found at any time point in children receiving bilateral implants simultaneously.

CONCLUSION:

Results suggest potential disruptions to binaural brainstem processing based on timing cues in children receiving a second cochlear implant after more than 2 years of unilateral implant use that persist through at least the first 9 months of bilateral implant use.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Potenciales Evocados Auditivos del Tronco Encefálico / Implantes Cocleares Límite: Child, preschool / Female / Humans / Infant / Male Idioma: En Revista: Otol Neurotol Asunto de la revista: NEUROLOGIA / OTORRINOLARINGOLOGIA Año: 2008 Tipo del documento: Article País de afiliación: Canadá

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Potenciales Evocados Auditivos del Tronco Encefálico / Implantes Cocleares Límite: Child, preschool / Female / Humans / Infant / Male Idioma: En Revista: Otol Neurotol Asunto de la revista: NEUROLOGIA / OTORRINOLARINGOLOGIA Año: 2008 Tipo del documento: Article País de afiliación: Canadá