The feasibility of epicardial echocardiography for measuring aortic valve area by the continuity equation.
Hilberath, Jan N; Shernan, Stanton K; Segal, Scott; Smith, Brian; Eltzschig, Holger K.
Afiliación
  • Hilberath JN; Department of Anesthesiology, Perioperative and Pain Medicine, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts, USA.
Anesth Analg ; 108(1): 17-22, 2009 Jan.
Article en En | MEDLINE | ID: mdl-19095825
ABSTRACT

BACKGROUND:

Measuring the aortic valve area (AVA) remains an important component of a comprehensive intraoperative echocardiographic examination in patients undergoing aortic valve surgery. Epicardial echocardiography (EE) represents an accessible alternative to transesophageal echocardiography (TEE), however, its agreement and correlation with other imaging modalities for measuring AVA has not been systematically validated.

METHODS:

EE was used in 85 patients undergoing cardiac surgery to measure AVA (AVA-EE) using the continuity equation. AVA-EE was compared to measurements obtained by intraoperative transesophageal echocardiography (AVA-TEE) in the same population. In a subset of patients, AVA-EE was also compared to AVA measurements from either preoperative transthoracic echocardiography (AVA-TTE) (n = 65) or cardiac catheterization (AVA-Cath) (n = 35) that were acquired within 4 wk before the date of surgery.

RESULTS:

Adequate trans-AV Doppler recordings were obtained in 94% of patients for AVA-TEE and 100% of patients for AVA-EE. EE measurements of AVA showed close agreement with TEE measurements (mean difference [bias] +/- 95% CI = -0.09 cm(2) +/- 0.18 cm(2), r(2) = 0.83, P < 0.0001). AVA-EE also agreed well with AVA-Cath (mean difference +/- 95% CI = -0.03 cm(2) +/- 0.12 cm(2), r(2) = 0.87, P < 0.0001) and AVA-TTE (mean difference +/- 95% CI = -0.06 cm(2) +/- 0.22 cm(2), r(2) = 0.81, P < 0.0001).

CONCLUSIONS:

EE measurements of AVA by the continuity equation show high agreement and closely correlate with established techniques of AVA assessment.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Válvula Aórtica / Pericardio / Ecocardiografía Doppler / Procedimientos Quirúrgicos Cardíacos / Modelos Cardiovasculares Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Anesth Analg Año: 2009 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Válvula Aórtica / Pericardio / Ecocardiografía Doppler / Procedimientos Quirúrgicos Cardíacos / Modelos Cardiovasculares Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Anesth Analg Año: 2009 Tipo del documento: Article País de afiliación: Estados Unidos