Titration calorimetry standards and the precision of isothermal titration calorimetry data.
Int J Mol Sci
; 10(6): 2752-2762, 2009 Jun 18.
Article
en En
| MEDLINE
| ID: mdl-19582227
Current Isothermal Titration Calorimetry (ITC) data in the literature have relatively high errors in the measured enthalpies of protein-ligand binding reactions. There is a need for universal validation standards for titration calorimeters. Several inorganic salt co-precipitation and buffer protonation reactions have been suggested as possible enthalpy standards. The performances of several commercial calorimeters, including the VP-ITC, ITC200, and Nano ITC-III, were validated using these suggested standard reactions.