Dynamic nuclear polarization system output volume reduction using inert fluids.
Peterson, Eric T; Gordon, Jeremy W; Erickson, Matthew G; Fain, Sean B; Rowland, Ian J.
Afiliación
  • Peterson ET; Department of Biomedical Engineering, University of Wisconsin, Madison, WI, USA. etpeterson@wisc.edu
J Magn Reson Imaging ; 33(4): 1003-8, 2011 Apr.
Article en En | MEDLINE | ID: mdl-21448970
PURPOSE: To present a method for significantly increasing the concentration of a hyperpolarized compound produced by a commercial dynamic nuclear polarization (DNP) polarizer, enabling the polarization process to be more suitable for preclinical applications. MATERIALS AND METHODS: Using a HyperSense DNP polarizer, we investigated the combined use of perfluorocarbon and water to warm and dissolve the hyperpolarized material from the polarization temperature of 1.4K to produce material at temperatures suitable for injection. RESULTS: By replacing 75% of the water in the dissolution volume with a chemically and biologically inert liquid that is immiscible with water, the injection volume can be reduced 4-fold. Rapid separation of the water and perfluorocarbon mixture enables the aqueous layer containing polarized material to be easily and rapidly collected. CONCLUSION: The approach provides a significantly increased concentration of compound in a volume for injection that is more appropriate for small animal studies. This is demonstrated for (13) C-labeled pyruvic acid and (13) C-labeled succinate, but may be applied to the majority of nuclei and compounds hyperpolarized by the DNP method.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Isótopos de Carbono / Diagnóstico por Imagen / Espectroscopía de Resonancia Magnética Tipo de estudio: Diagnostic_studies Límite: Animals Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2011 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Isótopos de Carbono / Diagnóstico por Imagen / Espectroscopía de Resonancia Magnética Tipo de estudio: Diagnostic_studies Límite: Animals Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2011 Tipo del documento: Article País de afiliación: Estados Unidos