A Novel Percutaneous Guide Wire (S-Wire) for Percutaneous Pedicle Screw Insertion: Its Development, Efficacy, and Safety.
Ishii, Ken; Kaneko, Yasuhito; Funao, Haruki; Ishihara, Shinichi; Shinohara, Akira; Nakanishi, Kazuo; Hikata, Tomohiro; Fujita, Nobuyuki; Iwanami, Akio; Hosogane, Naobumi; Watanabe, Kota; Nagura, Takeo; Nakamura, Masaya; Toyama, Yoshiaki; Matsumoto, Morio.
Afiliación
  • Ishii K; Keio University, Tokyo, Japan Society for Minimally Invasive Spine Stabilization, Nagoya, Aichi, Japan keni8888@z7.keio.jp.
  • Kaneko Y; Society for Minimally Invasive Spine Stabilization, Nagoya, Aichi, Japan Kawasaki Municipal Kawasaki Hospital, Kanagawa, Japan.
  • Funao H; Society for Minimally Invasive Spine Stabilization, Nagoya, Aichi, Japan Nerima General Hospital, Tokyo, Japan Johns Hopkins University, Baltimore, MD, USA.
  • Ishihara S; Keio University, Tokyo, Japan Society for Minimally Invasive Spine Stabilization, Nagoya, Aichi, Japan.
  • Shinohara A; Society for Minimally Invasive Spine Stabilization, Nagoya, Aichi, Japan Jikei University Kashiwa Hospital, Chiba, Japan.
  • Nakanishi K; Society for Minimally Invasive Spine Stabilization, Nagoya, Aichi, Japan Kawasaki Medical School, Okayama, Japan.
  • Hikata T; Keio University, Tokyo, Japan Society for Minimally Invasive Spine Stabilization, Nagoya, Aichi, Japan.
  • Fujita N; Keio University, Tokyo, Japan Society for Minimally Invasive Spine Stabilization, Nagoya, Aichi, Japan.
  • Iwanami A; Keio University, Tokyo, Japan Society for Minimally Invasive Spine Stabilization, Nagoya, Aichi, Japan.
  • Hosogane N; Keio University, Tokyo, Japan Society for Minimally Invasive Spine Stabilization, Nagoya, Aichi, Japan.
  • Watanabe K; Keio University, Tokyo, Japan.
  • Nagura T; Keio University, Tokyo, Japan.
  • Nakamura M; Keio University, Tokyo, Japan.
  • Toyama Y; Keio University, Tokyo, Japan.
  • Matsumoto M; Keio University, Tokyo, Japan.
Surg Innov ; 22(5): 469-73, 2015 Oct.
Article en En | MEDLINE | ID: mdl-25432881
ABSTRACT

PURPOSE:

Minimally invasive spine stabilization (MISt) procedures, including MIS-transforaminal lumbar interbody fusion (MIS-TLIF), rely on precise placement of percutaneous pedicle screws (PPS). Serious intraoperative complications associated with PPS placement include great vessel and bowel injuries due to the guide-wire's anterior migration and penetration through the anterior aspect of the vertebral body. To address this issue, we developed a novel percutaneous guide wire (S-wire) and compared the biomechanical characteristics of S-wire and conventional wire in cadaveric spines, and to evaluate the S-wire's efficacy and safety in a clinical trial.

METHODS:

The S-wire is hollow, with braided wires extending at one tip. We compared the push-out and penetration forces of the S-wire and conventional wire in fresh cadaveric lumbar spines, from L1 to L5.

RESULTS:

Push-out forces caused the braided tip of the S-wire to bend or spread, and thus to resist anterior migration. The mean push-out forces for the S-wire and conventional wire were 15.5 ± 1.9 and 5.7 ± 0.8 N, respectively (P < .0001); the mean penetration forces were 69.1 ± 4.2 and 37.1± 4.8 N, respectively (P < .0005). There was no wire breakage or anterior-wall penetration in a clinical trial of 922 S-wires; interestingly, the pull-out force increased in 780 (84.6%) S-wires after placement.

CONCLUSIONS:

The mean push-out and penetration forces for the S-wire were approximately 3 and 2 times greater than those of conventional wire, respectively. The S-wire effectively prevented guide-wire anterior migration and penetration of the anterior vertebral-body wall. The S-wire device should effectively improve the safety of MISt procedures, including MIS-TLIF and percutaneous kyphoplasty in selected patient with osteoporosis.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Hilos Ortopédicos / Procedimientos Ortopédicos / Tornillos Pediculares / Vértebras Lumbares Límite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: Surg Innov Año: 2015 Tipo del documento: Article País de afiliación: Japón

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Hilos Ortopédicos / Procedimientos Ortopédicos / Tornillos Pediculares / Vértebras Lumbares Límite: Aged / Aged80 / Female / Humans / Male Idioma: En Revista: Surg Innov Año: 2015 Tipo del documento: Article País de afiliación: Japón