Manual, semiautomated, and fully automated measurement of the aortic annulus for planning of transcatheter aortic valve replacement (TAVR/TAVI): analysis of interchangeability.
Lou, Junyang; Obuchowski, Nancy A; Krishnaswamy, Amar; Popovic, Zoran; Flamm, Scott D; Kapadia, Samir R; Svensson, Lars G; Bolen, Michael A; Desai, Milind Y; Halliburton, Sandra S; Tuzcu, E Murat; Schoenhagen, Paul.
Afiliación
  • Lou J; Heart & Vascular Institute, Cleveland Clinic, Cleveland, OH, USA.
  • Obuchowski NA; Cardiovascular Imaging, Imaging Institute, Cleveland Clinic, Desk J1-4, 9500 Euclid Avenue, Cleveland, OH, USA; Quantitative Health Sciences, Cleveland Clinic, Cleveland, OH, USA.
  • Krishnaswamy A; Heart & Vascular Institute, Cleveland Clinic, Cleveland, OH, USA.
  • Popovic Z; Heart & Vascular Institute, Cleveland Clinic, Cleveland, OH, USA; Cardiovascular Imaging, Imaging Institute, Cleveland Clinic, Desk J1-4, 9500 Euclid Avenue, Cleveland, OH, USA.
  • Flamm SD; Heart & Vascular Institute, Cleveland Clinic, Cleveland, OH, USA; Cardiovascular Imaging, Imaging Institute, Cleveland Clinic, Desk J1-4, 9500 Euclid Avenue, Cleveland, OH, USA.
  • Kapadia SR; Heart & Vascular Institute, Cleveland Clinic, Cleveland, OH, USA.
  • Svensson LG; Heart & Vascular Institute, Cleveland Clinic, Cleveland, OH, USA.
  • Bolen MA; Heart & Vascular Institute, Cleveland Clinic, Cleveland, OH, USA; Cardiovascular Imaging, Imaging Institute, Cleveland Clinic, Desk J1-4, 9500 Euclid Avenue, Cleveland, OH, USA.
  • Desai MY; Heart & Vascular Institute, Cleveland Clinic, Cleveland, OH, USA; Cardiovascular Imaging, Imaging Institute, Cleveland Clinic, Desk J1-4, 9500 Euclid Avenue, Cleveland, OH, USA.
  • Halliburton SS; Heart & Vascular Institute, Cleveland Clinic, Cleveland, OH, USA; Cardiovascular Imaging, Imaging Institute, Cleveland Clinic, Desk J1-4, 9500 Euclid Avenue, Cleveland, OH, USA.
  • Tuzcu EM; Heart & Vascular Institute, Cleveland Clinic, Cleveland, OH, USA.
  • Schoenhagen P; Heart & Vascular Institute, Cleveland Clinic, Cleveland, OH, USA; Cardiovascular Imaging, Imaging Institute, Cleveland Clinic, Desk J1-4, 9500 Euclid Avenue, Cleveland, OH, USA. Electronic address: schoenp1@ccf.org.
J Cardiovasc Comput Tomogr ; 9(1): 42-9, 2015.
Article en En | MEDLINE | ID: mdl-25533222
BACKGROUND: Preprocedural 3-dimensional CT imaging of the aortic annular plane plays a critical role for transcatheter aortic valve replacement (TAVR) planning; however, manual reconstructions are complex. Automated analysis software may improve reproducibility and agreement between readers but is incompletely validated. METHODS: In 110 TAVR patients (mean age, 81 years; 37% female) undergoing preprocedural multidetector CT, automated reconstruction of the aortic annular plane and planimetry of the annulus was performed with a prototype of now commercially available software (syngo.CT Cardiac Function-Valve Pilot; Siemens Healthcare, Erlangen, Germany). Fully automated, semiautomated, and manual annulus measurements were compared. Intrareader and inter-reader agreement, intermodality agreement, and interchangeability were analyzed. Finally, the impact of these measurements on recommended valve size was evaluated. RESULTS: Semiautomated analysis required major correction in 5 patients (4.5%). In the remaining 95.5%, only minor correction was performed. Mean manual annulus area was significantly smaller than fully automated results (P < .001 for both readers) but similar to semiautomated measurements (5.0 vs 5.4 vs 4.9 cm(2), respectively). The frequency of concordant recommendations for valve size increased if manual analysis was replaced with the semiautomated method (60% agreement was improved to 82.4%; 95% confidence interval for the difference [69.1%-83.4%]). CONCLUSIONS: Semiautomated aortic annulus analysis, with minor correction by the user, provides reliable results in the context of TAVR annulus evaluation.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Interpretación de Imagen Radiográfica Asistida por Computador / Tomografía Computarizada por Rayos X / Cirugía Asistida por Computador / Reemplazo de la Válvula Aórtica Transcatéter Tipo de estudio: Clinical_trials / Diagnostic_studies Límite: Adult / Aged / Aged80 / Humans / Middle aged Idioma: En Revista: J Cardiovasc Comput Tomogr Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / RADIOLOGIA Año: 2015 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Interpretación de Imagen Radiográfica Asistida por Computador / Tomografía Computarizada por Rayos X / Cirugía Asistida por Computador / Reemplazo de la Válvula Aórtica Transcatéter Tipo de estudio: Clinical_trials / Diagnostic_studies Límite: Adult / Aged / Aged80 / Humans / Middle aged Idioma: En Revista: J Cardiovasc Comput Tomogr Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / RADIOLOGIA Año: 2015 Tipo del documento: Article País de afiliación: Estados Unidos