Moving sum of number of positive patient result as a quality control tool.
Clin Chem Lab Med
; 55(11): 1709-1714, 2017 Oct 26.
Article
en En
| MEDLINE
| ID: mdl-28328525
BACKGROUND: Recently, the total prostate-specific antigen (PSA) assay used in a laboratory had a positive bias of 0.03 µg/L, which went undetected. Consequently, a number of post-prostatectomy patients with previously undetectable PSA concentrations (defined as <0.03 µg/L in that laboratory) were being reported as having detectable PSA, which suggested poorer prognosis according to clinical guidelines. METHODS: Through numerical simulations, we explored (1) how a small bias may evade the detection of routine quality control (QC) procedures with specific reference to the concentration of the QC material, (2) whether the use of 'average of normals' approach may detect such a small bias, and (3) describe the use of moving sum of number of patient results with detectable PSA as an adjunct QC procedure. RESULTS: The lowest QC level (0.86 µg/L) available from a commercial kit had poor probability (<10%) of a bias of 0.03 µg/L regardless of QC rule (i.e. 1:2S, 2:2S, 1:3S, 4:1S) used. The average number of patient results affected before error detection (ANPed) was high when using the average of normals approach due to the relatively wide control limits. By contrast, the ANPed was significantly lower for the moving sum of number of patient results with a detectable PSA approach. CONCLUSIONS: Laboratory practitioners should ensure their QC strategy can detect small but critical bias, and may require supplementation of ultra-low QC levels that are not covered by commercial kits with in-house preparations. The use of moving sum of number of patient results with a detectable result is a helpful adjunct QC tool.
Palabras clave
Texto completo:
1
Banco de datos:
MEDLINE
Asunto principal:
Antígeno Prostático Específico
/
Pruebas de Química Clínica
Tipo de estudio:
Diagnostic_studies
/
Guideline
/
Prognostic_studies
Límite:
Humans
/
Male
Idioma:
En
Revista:
Clin Chem Lab Med
Asunto de la revista:
QUIMICA CLINICA
/
TECNICAS E PROCEDIMENTOS DE LABORATORIO
Año:
2017
Tipo del documento:
Article