Your browser doesn't support javascript.
loading
Improved diagnosis of the number of stenosed coronary artery vessels by segmentation with scatter and photo-peak window data for attenuation correction in myocardial perfusion SPECT.
Yamauchi, Yohei; Kanzaki, Yumiko; Hayashi, Masuo; Arai, Mami; Morita, Hideaki; Komori, Tsuyoshi; Hoshiga, Masaaki; Ishizaka, Nobukazu.
Afiliación
  • Yamauchi Y; Department of Cardiology, Osaka Medical College, 2-7, Daigaku-machi, Takatsuki, 569-8686, Japan.
  • Kanzaki Y; Department of Cardiology, Osaka Medical College, 2-7, Daigaku-machi, Takatsuki, 569-8686, Japan. in3089@osaka-med.ac.jp.
  • Hayashi M; Department of Radiology, Osaka Medical College, Takatsuki, Japan.
  • Arai M; Department of Radiology, Osaka Medical College, Takatsuki, Japan.
  • Morita H; Department of Cardiology, Osaka Medical College, 2-7, Daigaku-machi, Takatsuki, 569-8686, Japan.
  • Komori T; Department of Radiology, Osaka Medical College, Takatsuki, Japan.
  • Hoshiga M; Department of Cardiology, Osaka Medical College, 2-7, Daigaku-machi, Takatsuki, 569-8686, Japan.
  • Ishizaka N; Department of Cardiology, Osaka Medical College, 2-7, Daigaku-machi, Takatsuki, 569-8686, Japan.
J Nucl Cardiol ; 26(2): 574-581, 2019 04.
Article en En | MEDLINE | ID: mdl-28905206
ABSTRACT

BACKGROUND:

Attenuation correction using segmentation of scatter and photo-peak window data (SSPAC) enables an evaluation of the attenuation map in a patient-specific manner without additional radiation exposure. We compared the accuracy of SSPAC and non-corrected myocardial perfusion scintigraphy methods for diagnosing the number of stenosed coronary artery vessels. METHODS AND

RESULTS:

We retrospectively reviewed the data from 183 consecutive patients who underwent 99mTc-tetrofosmin stress/rest SPECT examination and a coronary angiography within 3 months. The MPS images were reconstructed with and without SSPAC attenuation correction. We examined the accuracy of the quantitative interpretation using summed differential score in the detection of coronary artery disease (CAD). The attenuation maps were successfully determined in 179 of 183 patients (98%). In terms of the vessel-based diagnostic ability, sensitivity, specificity, positive predictive and negative predictive values of the SSPAC and non-correction methods for diagnosing CAD in individual coronary territories were 77%*, 89%, 74%*, and 90%* vs 51%, 87%, 62%, and 82%, respectively (*P < .05). In 35 patients with multi-vessel CAD, those values were 78%*, 81%, 93%, and 55%* vs 49%, 81%, 89%, and 34%, respectively (*P < .05; AUC 0.82 vs 0.62, P < .05).

CONCLUSION:

SSPAC-corrected SPECT myocardial perfusion images exhibit improved accuracy in the detection of the number of stenosed coronary artery vessels, even in patients with multi-vessel CAD.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Tomografía Computarizada de Emisión de Fotón Único / Angiografía Coronaria / Vasos Coronarios / Imagen de Perfusión Miocárdica Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Japón

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Tomografía Computarizada de Emisión de Fotón Único / Angiografía Coronaria / Vasos Coronarios / Imagen de Perfusión Miocárdica Tipo de estudio: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Nucl Cardiol Asunto de la revista: CARDIOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Japón