The usefulness of mobile insulator sheets for the optimisation of deep heating area for regional hyperthermia using a capacitively coupled heating method: phantom, simulation and clinical prospective studies.
Tomura, Kyosuke; Ohguri, Takayuki; Mulder, Hendrik Thijmen; Murakami, Motohiro; Nakahara, Sota; Yahara, Katsuya; Korogi, Yukunori.
Afiliación
  • Tomura K; a Department of Radiology , University of Occupational and Environmental Health , Kitakyushu , Japan.
  • Ohguri T; a Department of Radiology , University of Occupational and Environmental Health , Kitakyushu , Japan.
  • Mulder HT; b Department of Radiation Oncology, Hyperthermia unit , Erasmus MC Cancer Institute , Rotterdam , the Netherlands.
  • Murakami M; c Department of Medical Electronics , University Hospital of Occupational and Environmental Health , Kitakyushu , Japan.
  • Nakahara S; a Department of Radiology , University of Occupational and Environmental Health , Kitakyushu , Japan.
  • Yahara K; a Department of Radiology , University of Occupational and Environmental Health , Kitakyushu , Japan.
  • Korogi Y; a Department of Radiology , University of Occupational and Environmental Health , Kitakyushu , Japan.
Int J Hyperthermia ; 34(7): 1092-1103, 2018 11.
Article en En | MEDLINE | ID: mdl-29108446
PURPOSE: To evaluate the feasibility and efficacy of deep regional hyperthermia with the use of mobile insulator sheets in a capacitively coupled heating device. MATERIALS AND METHODS: The heat was applied using an 8-MHz radiofrequency-capacitive device. The insulator sheet was inserted between the regular bolus and cooled overlay bolus in each of upper and lower side of the electrode. Several settings using the insulator sheets were investigated in an experimental study using an agar phantom to evaluate the temperature distributions. The specific absorption rate (SAR) distributions in several organs were also computed for the three-dimensional patient model. In a clinical prospective study, a total of five heating sessions were scheduled for the pelvic tumours, to assess the thermal parameters. The conventional setting was used during the first, third and fifth treatment sessions, and insulator sheets were used during the second and fourth treatment sessions. RESULTS: In the phantom study, the higher heating area improved towards the centre when the mobile insulator sheets were used. The subcutaneous fat/target ratios for the averaged SARs in the setting with the mobile insulator (median, 2.5) were significantly improved compared with those in the conventional setting (median, 3.4). In the clinical study, the thermal dose parameters of CEM43°CT90 in the sessions with the mobile insulator sheets (median, 1.9 min) were significantly better than those in the sessions using a conventional setting (median, 1.0 min). CONCLUSIONS: Our novel heating method using mobile insulator sheets was thus found to improve the thermal dose parameters. Further investigations are expected.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Salicilatos / Fantasmas de Imagen / Hipertermia Inducida / Niacina Tipo de estudio: Observational_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Int J Hyperthermia Asunto de la revista: NEOPLASIAS / TERAPEUTICA Año: 2018 Tipo del documento: Article País de afiliación: Japón

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Salicilatos / Fantasmas de Imagen / Hipertermia Inducida / Niacina Tipo de estudio: Observational_studies / Prognostic_studies Límite: Humans Idioma: En Revista: Int J Hyperthermia Asunto de la revista: NEOPLASIAS / TERAPEUTICA Año: 2018 Tipo del documento: Article País de afiliación: Japón