[Seven Cases of Troubleshooting and Analysis on VARIAN IX Medical Linear Accelerator].
Zhongguo Yi Liao Qi Xie Za Zhi
; 40(5): 388-90, 2016 Sep.
Article
en Zh
| MEDLINE
| ID: mdl-29792641
ABSTRACT
This paper briefly introduced seven cases of malfunctions occurred in the VARIAN IX linear accelerator during operation, i.e., control board malfunction of electronic gun (GFIL interlock), modulator malfunction (HVCB interlock), energy programming board malfunction (EXQ1), energy conversion potentiometer malfunction (CARR interlock), MLC malfunction, thermostat valve malfunction (PUMP interlock) and ionization chamber malfunction (ION interlock). Moreover, malfunction cases analyzed and troubleshooting methods presented in this paper can provide reference for colleagues.
Buscar en Google
Banco de datos:
MEDLINE
Asunto principal:
Aceleradores de Partículas
Idioma:
Zh
Revista:
Zhongguo Yi Liao Qi Xie Za Zhi
Año:
2016
Tipo del documento:
Article