Dual-energy CT and ceramic or titanium prostheses material reduce CT artifacts and provide superior image quality of total knee arthroplasty.
Kasparek, Maximilian F; Töpker, Michael; Lazar, Mathias; Weber, Michael; Kasparek, Michael; Mang, Thomas; Apfaltrer, Paul; Kubista, Bernd; Windhager, Reinhard; Ringl, Helmut.
Afiliación
  • Kasparek MF; Department of Orthopaedics and Trauma Surgery, Vienna General Hospital, Medical University of Vienna, Waehringer Guertel 18-20, 1090, Vienna, Austria.
  • Töpker M; Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090, Vienna, Austria.
  • Lazar M; Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090, Vienna, Austria.
  • Weber M; Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090, Vienna, Austria.
  • Kasparek M; Department of Orthopaedic Surgery, Orthopaedic Hospital Speising, Speisinger Str. 109, 1130, Vienna, Austria.
  • Mang T; Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090, Vienna, Austria.
  • Apfaltrer P; Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090, Vienna, Austria.
  • Kubista B; Department of Orthopaedics and Trauma Surgery, Vienna General Hospital, Medical University of Vienna, Waehringer Guertel 18-20, 1090, Vienna, Austria.
  • Windhager R; Department of Orthopaedics and Trauma Surgery, Vienna General Hospital, Medical University of Vienna, Waehringer Guertel 18-20, 1090, Vienna, Austria.
  • Ringl H; Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Waehringer Guertel 18-20, 1090, Vienna, Austria. helmut.ringl@meduniwien.ac.at.
Knee Surg Sports Traumatol Arthrosc ; 27(5): 1552-1561, 2019 May.
Article en En | MEDLINE | ID: mdl-29881885
ABSTRACT

PURPOSE:

To evaluate the influence of different scan parameters for single-energy CT and dual-energy CT, as well as the impact of different material used in a TKA prosthesis on image quality and the extent of metal artifacts.

METHODS:

Eight pairs of TKA prostheses from different vendors were examined in a phantom set-up. Each pair consisted of a conventional CoCr prosthesis and the corresponding anti-allergic prosthesis (full titanium, ceramic, or ceramic-coated) from the same vendor. Nine different (seven dual-energy CT and two single-energy CT) scan protocols with different characteristics were used to determine the most suitable CT protocol for TKA imaging. Quantitative image analysis included assessment of blooming artifacts (metal implants appear thicker on CT than they are, given as virtual growth in mm in this paper) and streak artifacts (thick dark lines around metal). Qualitative image analysis was used to investigate the bone-prosthesis interface.

RESULTS:

The full titanium prosthesis and full ceramic knee showed significantly fewer blooming artifacts compared to the standard CoCr prosthesis (mean virtual growth 0.6-2.2 mm compared to 2.9-4.6 mm, p < 0.001). Dual-energy CT protocols showed less blooming (range 3.3-3.8 mm) compared to single-energy protocols (4.6-5.5 mm). The full titanium and full ceramic prostheses showed significantly fewer streak artifacts (mean standard deviation 77-86 Hounsfield unit (HU)) compared to the standard CoCr prosthesis (277-334 HU, p < 0.001). All dual-energy CT protocols had fewer metal streak artifacts (215-296 HU compared to single-energy CT protocols (392-497 HU)). Full titanium and ceramic prostheses were ranked superior with regard to the image quality at the bone/prosthesis interface compared to a standard CoCr prosthesis, and all dual-energy CT protocols were ranked better than single-energy protocols.

CONCLUSIONS:

Dual-energy CT and ceramic or titanium prostheses reduce CT artifacts and provide superior image quality of total knee arthroplasty at the bone/prosthesis interface. These findings support the use of dual-energy CT as a solid imaging base for clinical decision-making and the use of full-titanium or ceramic prostheses to allow for better CT visualization of the bone-prosthesis interface.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Titanio / Interpretación de Imagen Radiográfica Asistida por Computador / Artefactos / Artroplastia de Reemplazo de Rodilla / Metales Tipo de estudio: Guideline / Prognostic_studies / Qualitative_research Límite: Humans Idioma: En Revista: Knee Surg Sports Traumatol Arthrosc Asunto de la revista: MEDICINA ESPORTIVA / TRAUMATOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Austria

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Titanio / Interpretación de Imagen Radiográfica Asistida por Computador / Artefactos / Artroplastia de Reemplazo de Rodilla / Metales Tipo de estudio: Guideline / Prognostic_studies / Qualitative_research Límite: Humans Idioma: En Revista: Knee Surg Sports Traumatol Arthrosc Asunto de la revista: MEDICINA ESPORTIVA / TRAUMATOLOGIA Año: 2019 Tipo del documento: Article País de afiliación: Austria