Intra- and inter-operator reproducibility of automated cloud-based carotid lumen diameter ultrasound measurement.
Saba, Luca; Banchhor, Sumit K; Araki, Tadashi; Viskovic, Klaudija; Londhe, Narendra D; Laird, John R; Suri, Harman S; Suri, Jasjit S.
Afiliación
  • Saba L; Department of Radiology, University of Cagliari, Italy.
  • Banchhor SK; Department of Electrical Engineering, NIT Raipur, Chhattisgarh, India.
  • Araki T; Division of Cardiovascular Medicine, Toho University Ohashi Medical Center, Tokyo, Japan.
  • Viskovic K; Department of Radiology and Ultrasound, University Hospital for Infectious Disease, Zagreb, Croatia.
  • Londhe ND; Department of Electrical Engineering, NIT Raipur, Chhattisgarh, India.
  • Laird JR; UC Davis Vascular Centre, University of California, Davis, CA, USA.
  • Suri HS; Monitoring and Diagnostic Division, AtheroPointÔ, Roseville, CA, USA, USA.
  • Suri JS; Monitoring and Diagnostic Division, AtheroPointÔ, Roseville, CA, USA, USA; Point-of-Care Devices, Global Biomedical Technologies, Inc., Roseville, CA, USA; Department of Electrical Engineering, University of Idaho (Affl.), ID, USA. Electronic address: Jasjit.Suri@atheropoint.com.
Indian Heart J ; 70(5): 649-664, 2018.
Article en En | MEDLINE | ID: mdl-30392503
ABSTRACT

BACKGROUND:

Common carotid artery lumen diameter (LD) ultrasound measurement systems are either manual or semi-automated and lack reproducibility and variability studies. This pilot study presents an automated and cloud-based LD measurements software system (AtheroCloud) and evaluates its (i) intra/inter-operator reproducibility and (ii) intra/inter-observer variability.

METHODS:

100 patients (83M, mean age 68±11years), IRB approved, consisted of L/R CCA artery (200 ultrasound images), acquired using a 7.5-MHz linear transducer. The intra/inter-operator reproducibility was verified using three operator's readings. Near-wall and far carotid wall borders were manually traced by two observers for intra/inter-observer variability analysis.

RESULTS:

The mean coefficient of correlation (CC) for intra- and inter-operator reproducibility between all the three automated reading pairs were 0.99 (P<0.0001) and 0.97 (P<0.0001), respectively. The mean CC for intra- and inter-observer variability between both the manual reading pairs were 0.98 (P<0.0001) and 0.98 (P<0.0001), respectively. The Figure-of-Merit between the mean of the three automated readings against the four manuals were 98.32%, 99.50%, 98.94% and 98.49%, respectively.

CONCLUSIONS:

The AtheroCloud LD measurement system showed high intra/inter-operator reproducibility hence can be adapted for vascular screening mode or pharmaceutical clinical trial mode.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Estenosis Carotídea / Arteria Carótida Común / Ultrasonografía Doppler / Grosor Intima-Media Carotídeo / Nube Computacional Tipo de estudio: Guideline / Observational_studies / Prognostic_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Indian Heart J Año: 2018 Tipo del documento: Article País de afiliación: Italia

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Estenosis Carotídea / Arteria Carótida Común / Ultrasonografía Doppler / Grosor Intima-Media Carotídeo / Nube Computacional Tipo de estudio: Guideline / Observational_studies / Prognostic_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Indian Heart J Año: 2018 Tipo del documento: Article País de afiliación: Italia