Evaluation of Nova StatStrip and FreeStyle Precision Pro blood ketone tests using 3-hydroxybutyrate doped samples.
Helmersson-Karlqvist, Johanna; Höög Hammarström, Kristina; Palmberg, Kicki; Backman-Johansson, Carolina.
Afiliación
  • Helmersson-Karlqvist J; Department of Medical Sciences/Clinical Chemistry, Uppsala University, Uppsala, Sweden.
  • Höög Hammarström K; Centre for Inherited Metabolic Diseases, Karolinska University Hospital, Stockholm, Sweden.
  • Palmberg K; Department of Medical Sciences/Clinical Chemistry, Uppsala University, Uppsala, Sweden.
  • Backman-Johansson C; Centre for Inherited Metabolic Diseases, Karolinska University Hospital, Stockholm, Sweden.
J Clin Lab Anal ; 33(4): e22851, 2019 May.
Article en En | MEDLINE | ID: mdl-30811634
BACKGROUND: The most clinically useful blood ketone in the diagnosis, management, and recovery of diabetes ketoacidosis in both adults and children is 3-hydroxybutyrate. In the absence of laboratory routine methods, several point-of-care methods are in use, but very few clinical evaluations are published. METHODS: This study evaluates linearity and reproducibility of two handheld point-of-care meters for blood 3-hydroxybutyrate measurement for hospital use, Nova StatStrip, and FreeStyle Precision Pro. Whole blood from healthy volunteers was spiked with different concentrations of a 3-hydroxybutyrate solution and tested on the point-of-care instruments. The results were compared with plasma 3-hydroxybutyrate that was analyzed with a laboratory enzymatic end point spectrophotometric reference method. RESULTS: Blood 3-hydroxybutyrate on StatStrip was linear with the reference method up to approximately 4 mmol/L, and FreeStyle was linear up to 6 mmol/L. At higher concentrations, the point-of-care instruments gave falsely too low results, especially the StatStrip meter. The FreeStyle meter had better precision and less bias than StatStrip. CONCLUSION: In the acute setting of diabetes ketoacidosis, blood 3-hydroxybutyrate in the higher ranges should be interpreted with caution as the point-of-care meters are less accurate there.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Análisis Químico de la Sangre / Ácido 3-Hidroxibutírico Tipo de estudio: Evaluation_studies Límite: Humans Idioma: En Revista: J Clin Lab Anal Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2019 Tipo del documento: Article País de afiliación: Suecia

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Análisis Químico de la Sangre / Ácido 3-Hidroxibutírico Tipo de estudio: Evaluation_studies Límite: Humans Idioma: En Revista: J Clin Lab Anal Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2019 Tipo del documento: Article País de afiliación: Suecia